149-H/149-HS/212-J/412-J Microvac
Description
149H Microvac Rotary Piston Pump 230/460 V, 3 Ø, 60 Hz with 230/460 V Coil
149H Microvac Rotary Piston Pump 400 V, 3 Ø, 60 Hz with 380 V Coil
149H Microvac Rotary Piston Pump 400 V, 3 Ø , 60 Hz with 415 V Coil
149H Microvac Rotary Piston Pump 400 V, 3 Ø, 60 Hz with 380 V Coil with Water Miser
149H Microvac Rotary Piston Pump 400 V, 3 Ø, 60 Hz with 415 V Coil with Water Miser
149HS Microvac Rotary Piston Pump 230/460 V, 3 Ø, 60 Hz with 230/460 V Coil
212J Microvac Rotary Piston Pump 230/460 V, 3 Ø, 60 Hz with 230/460 V Coil
212J Microvac Rotary Piston Pump 400 V, 3 Ø, 60 Hz with 380 V Coil
212J Microvac Rotary Piston Pump 400 V, 3 Ø, 60 Hz with 415 V Coil
212J Microvac Rotary Piston Pump 400 V, 3 Ø, 60 Hz with 380 V Coil with Water Miser
212J Microvac Rotary Piston Pump 400 V, 3 Ø, 60 Hz with 415 V Coil with Water Miser
412J Microvac Rotary Piston Pump 230/460 V, 3 Ø, 60 Hz with 230/460 V Coil
412J Microvac Rotary Piston Pump 400 V, 3 Ø, 50 Hz with 380 V Coil
412J Microvac Rotary Piston Pump 400 V, 3 Ø, 50 Hz with 415 V Coil
412J Microvac Rotary Piston Pump 400 V, 3 Ø, 50 Hz with 380 V Coil with Water Miser
412J Microvac Rotary Piston Pump 40 0V, 3 Ø, 50 Hz with 415 V Coil with Water Miser
Instruction Manual
S149-01-880
Issue F Original
®
Pumps
Item Number
900-149-011
900149011501
900149011502
900149011503
900149011504
930-149-011
900-212-014
900212014501
900212014502
900212014503
900212014504
900-412-014
900412014501
900412014502
900412014503
900412014504

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Microvac 149-HS and is the answer not in the manual?

Questions and answers

Summary of Contents for Edwards Microvac 149-HS

  • Page 1 S149-01-880 Issue F Original Instruction Manual ® 149-H/149-HS/212-J/412-J Microvac Pumps Description Item Number 149H Microvac Rotary Piston Pump 230/460 V, 3 Ø, 60 Hz with 230/460 V Coil 900-149-011 149H Microvac Rotary Piston Pump 400 V, 3 Ø, 60 Hz with 380 V Coil 900149011501 149H Microvac Rotary Piston Pump 400 V, 3 Ø...
  • Page 2 Declaration of Conformity Edwards, Manor Royal, Crawley, West Sussex, RH10 9LW, UK declare under our sole responsibility, as manufacturer and person within the EU authorised to assemble the technical file, that the product(s) 900 YYYYYY 50 Z Pump Model Voltage/TCV Option...
  • Page 3: Table Of Contents

    Drive belt tension .......................30 Gas ballast check valve replacement ................30 5.10 Fitment or replacement of the motor (412J models only) ............31 © Edwards Limited 2011. All rights reserved. Page i Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 4 Electrical schematic ....................19 Gas ballast setting ......................24 Typical vacuum system piping ..................28 Collar bolt positions and motor frame sizes ................ 31 Page ii © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 5 Publication title Publication number Vacuum pump and vacuum system safety P300-20-000 Trademark credits ® Fomblin is a registered trademark of Ausimont SpA. © Edwards Limited 2011. All rights reserved. Page iii Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 6 S149-01-880 Issue F This page has been intentionally left blank. Page iv © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 7: Introduction

    Warning - risk of electric shock. documentation. Warning - hot surfaces. Warning - risk of explosion Warning - use protective equipment Warning - moving parts © Edwards Limited 2011. All rights reserved. Page 1 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 8: Atex Directive Implications

    212-J, and 7.6 LPM (2 GPM) for the 412-J maximum is required for efficient pump performance. Internal pump water pressure should not exceed 35 psig (Refer to Section 3.8 for information). Page 2 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 9: 149-H/149-Hs Pump General Arrangement

    12. 1/2 inch vacuum access port 13. Oil drain valve 14. Oil flow indicator 15. Vacuum exhaust connection 16. Belt guard 17. Solenoid valve © Edwards Limited 2011. All rights reserved. Page 3 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 10: 212-J Pump General Arrangement

    13. Oil level sight glass 14. Oil fill port 15. Motor 16. System lift eye-bolt 17. Reservoir cover 18. Oil drain valve Page 4 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 11: 412-J Pump General Arrangement

    17. Gas ballast purge connection 18. Gas ballast adjustment 19. 1/4 inch vacuum access port 20. Belt guard 21. Vacuum inlet connection © Edwards Limited 2011. All rights reserved. Page 5 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 12: Electrical System

    A 1/4 inch oil drain access port has been added to provide a means of removing oil that pools above the valve decks (Refer to Figure 4 Item 7). Page 6 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 13: Technical Data

    26-1/4 x 24 40-1/4 x 25-5/8 597 x 511 597 x 511 667 x 610 1022 x 651 Cooling water consumption (maximum) © Edwards Limited 2011. All rights reserved. Page 7 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 14: Materials In Contact With Process Gases

    Items ending with 501 or 503 utilize a 380 VAC, 50 Hz solenoid. Items ending with 502 or 504 utilize a 415 VAC, 50 Hz solenoid. Figure 5 - 149-H/149-HS pump dimensions Page 8 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 15: 212-J Pump Dimensions

    3/50 Hz 3/50 Hz 11.0 Note: Refer to motor rating nameplate for weight, efficiency, motor speed, and full load amperage values. © Edwards Limited 2011. All rights reserved. Page 9 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 16: 412-J Pump Dimensions

    215T Insulation class Electrical supply 230/460 230/460 230/460 230/460 Service factor 1.15 1.15 1.15 1.15 Duty Continuous Continuous Continuous Continuous Mounting Page 10 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 17: Pump Speed Curves

    Refer to motor rating nameplate for weight, efficiency, motor speed, and full load amperage values. Figure 8 - Pump speed curves © Edwards Limited 2011. All rights reserved. Page 11 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 18 S149-01-880 Issue F This page has been intentionally left blank. Page 12 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 19: Installation

    Ensure that the Microvac pump is suitable for the application. Refer to the Edwards guidelines on vacuum pump and vacuum system safety (see the associated publications at the end of the Contents list at the front of this manual).
  • Page 20: System Design Considerations

    If the unit will be stored for a period of time before commissioning, follow the same procedure. Page 14 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 21: Inlet Piping

    (See Figure 10). A full-opening high vacuum valve is recommended to facilitate start-up and for checking the pump blank off pressure. © Edwards Limited 2011. All rights reserved. Page 15 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 22: Vacuum Access Port Locations

    Tygon should be used for a flexible connection. The gauge should be located approximately 610 mm (24.0 inch) above the port to prevent oil from entering the gauge. Page 16 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 23: Exhaust Piping

    138 mbar (2 psig). The pumps emit tiny oil droplets into the exhaust of the pump. Edwards recommends the use of exhaust filters to prevent these droplets from accumulating in the exhaust piping of the system or where the piping terminates. The oil and condensables will coalesce inside the piping and return to the pump reservoir or be collected inside the filter housing.
  • Page 24: Electrical Connections

    Connect the motor so that the pump shaft rotates clockwise when viewed from the drive end. See Section 4.1 pre-start check procedures. Page 18 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 25: Electrical Schematic

    S149-01-880 Issue F Figure 11 - Electrical schematic Fuse Motor contactor M Aux In-line filter Overload Solenoid valve © Edwards Limited 2011. All rights reserved. Page 19 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 26: Cooling

    71°C (140-160°F). Oil temperature kits are available that automatically control the water flow to maintain the proper oil temperature (Contact Edwards for additional information). If the pump is located outside a building and subjected to freezing temperatures, the pump water jacket should be drained or filled with anti-freeze.
  • Page 27: Initial Oil Fill

    212-J; and, 45 liters (12 gallons) for the 412-J. Before connecting the suction manifold, slowly rotate the pump through two (2) revolutions. This will distribute the oil throughout the pump interior. © Edwards Limited 2011. All rights reserved. Page 21 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 28 S149-01-880 Issue F This page has been intentionally left blank. Page 22 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 29: Operation

    2. Ensure the equipment being evacuated is properly cleaned and all vacuum breaks are closed. 3. Depress the START button and check the solenoid valve for proper operation. 4. Open the pump inlet valve. © Edwards Limited 2011. All rights reserved. Page 23 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 30: Gas Ballast Operation

    If desired, a solenoid valve attached to the gas ballast inlet and electrically connected to the motor leads can be used to turn off the gas ballast automatically on pump shutdown. Contact Edwards for additional information.
  • Page 31: Pump Stop

    Note: Edwards offers a closed type oil mist separator that will eliminate exhaust oil smog but still requires exhausting to outside. Contact a Edwards representative for information. If the pump has been shut down for an extended period of time, it will be necessary to turn the pump over at least two (2) revolutions by momentarily jogging before starting to insure the free movement of parts.
  • Page 32 S149-01-880 Issue F This page has been intentionally left blank. Page 26 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 33: Maintenance

    (590 °F) and above. The pump may have overheated if it was misused or if it was in a fire. Health and Safety Data sheets for fluorinated materials used in the pump are available on request: contact Edwards or the supplier.
  • Page 34: Maintenance Plan

    12 months Clean the motor and drive 12 months Inspect taper lock bushings on pulleys After break in and 12 months Page 28 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 35: Oil Change

    Review the process contaminants and select proper personal protective equipment (PPE) to protect against media contact. Using the drain valve, drain the oil into a suitable container for disposal. Use only Edwards recommended oils when refilling the pump. Refer to Section 2 for fluid type and quantities.
  • Page 36: Belt Maintenance

    Use an O-ring pick or suitable tool to pull out the gas ballast check valve. Replace the valve if the spring is damaged or broken. Reinstall the stem valve. Check the pump and system for potential hazardous conditions. Restore energy sources. Page 30 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 37: Fitment Or Replacement Of The Motor (412J Models Only)

    900-412-014 NEMA 254T 60 Hz 5 inch -4.0 º B105 (L=2720) Figure 14 - Collar bolt positions and motor frame sizes © Edwards Limited 2011. All rights reserved. Page 31 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 38: Fault Finding

    Page 32 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 39 Atmospheric air prevents reversing direction of pump piston on pump shutdown. Oil will also be prevented from back-streaming into inlet piping. © Edwards Limited 2011. All rights reserved. Page 33 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 40 S149-01-880 Issue F This page has been intentionally left blank. Page 34 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 41: Storage And Disposal

    Particular care should be taken with components and waste oil which have been contaminated with dangerous process substances. Do not incinerate fluoroelastomer seals and O-rings. © Edwards Limited 2011. All rights reserved. Page 35 Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 42 S149-01-880 Issue F This page has been intentionally left blank. Page 36 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 43: Spares And Accessories

    Equipment which has been serviced, repaired or rebuilt is returned with a full warranty. The local Service Center can also provide Edwards engineers to support on-site maintenance, service or repair of your equipment.
  • Page 44 S149-01-880 Issue F This page has been intentionally left blank. Page 38 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.
  • Page 45: Pfpe-Prepared Microvac Pumps

    Edwards. Fomblin oil has different properties from other pump oils, therefore: If filling the PFPE-prepared Microvac pump with Fomblin oil, Edwards recommends to check for oil leaks, particularly around the shaft seals. If an oil leak is detected, contact the supplier or Edwards for advice.
  • Page 46 S149-01-880 Issue F This page has been intentionally left blank. Page 40 © Edwards Limited 2011. All rights reserved. Edwards and the Edwards logo are trademarks of Edwards Limited.