Commissioning
4.4 Establishing a mains connection
20
NOTICE!
Use of impermissible mains voltage or mains fre-
quency
Device damage
Compare the type plate with the available mains
voltage and mains frequency.
Take the following information into account:
Only use the supplied power cable for the power supply.
The mains plug disconnects the device from the power supply.
The mains plug must be easy to identify and access.
Only connect the device to sockets with a protective earth con-
ductor (PE).
The devices must be secured with a 16 A circuit breaker during
installation.
Exception: Devices with 13 A UK plugs.