Technical Specifications - Interacoustics AS608 Instructions For Use Manual

Screening audiometer
Hide thumbs Also See for AS608:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
The technical specifications provided here covers the general aspects of the instrument, while more
specific details are found in the Operation Manual in English.
Standards:
Audiometer
Safety
EMC
Calibration:
ISO 389-1 (DD45)
ISO 389-2 (EAR-Tone 3A and 5A)
PTB Report 1.61-64/04 (HDA280).
PTB Report 1.61-4039503/09 (DD45)
Medical CE-mark:
The CE-mark indicates that Interacoustics A/S meets the requirements of Annex II of the
Medical Device Directive 93/42/EEC. Approval of the quality system is made by TÜV –
identification no. 0123.
Frequencies and Maximum Intensities;
Freq.
Hz.
dB HL
125
250
500
750
1000
1500
2000
3000
4000
6000
8000
Inputs:
Tone, Warble Tone.
Outputs:
Left AC, Right AC.
Attenuator:
-10 to 100 dB HL in 5 dB steps.
Tone Presentation:
Manual or reverse (chosen in Setup Menu).
Multiple pulses 250 or 500msec (chosen in Setup Menu). On/off.
Talk Forward
(AS608e):
Auto Threshold
(AS608e):
AS608/AS608e Instruction for Use - English
Date: 2011-01-27

Technical Specifications

:
IEC 60645-1/ANSI S3.6, Type 4
:
IEC 60601-1
:
IEC 60601-1-2
AC
70
90
100
100
100
100
100
100
100
100
90
Built in talk forward microphone. 0-110dB SPL. Adjustable.
Patient controlled Hughson Westlake procedure according to ISO 8253-1.
Page 5/7