2. Hang the pump by the foot end of the bed,
or place it on a flat sturdy surface. Place the
power cord so there is no risk of stumbling
over it, run it over with the bed wheels, or
getting it jammed when raising or lowering
the bed.
4. Check to make sure that the power is set to '0' (off). Plug in the power cord into an
approved outlet (220 V).
NOTE!
• Make up the bed with sheets to increase user comfort.
• Check that the CPR-valve is closed.
3.2 Start-up and settings
NOTE!
Always read the user manual prior to use.
1. Start the pump by pressing '1' (on), on the side of the pump. The green light is lit.
2. The pump starts filling the mattress with air. It takes approximately 15-20 minutes.
3.3 Comfort settings
The air pressure in the mattress can be adjusted according to the
patient´s desire for comfort. Always consult the prescriber or
other authorised personnel for the appropriate setting.
Adjust the pressure with the comfort setting knob according to
the table on the pump.
NOTE!
• When the head end of the bed is raised, the pressure should be increased at least one
(1) step.
• For amputated users, when only part of the mattress is under load, the pressure should
be increased by 1-2 steps.
• Recommended users weight is up to 200 kg.
NOTE! A pump bag is
available as an option.
3. Connect the mattress'
air hoses to the side of
the pump. If the pump
and hoses are equipped
with quick connectors,
an audible "click" will be
heard when these have been connected correctly.