Instructions for use 2020-12-08 CuroCell S.A.M. ® Constant low pressure reactive air mattress system Instructions for use item number: 95-001298...
We would be very grateful for all information involving accidents or incidents relating to our products, by reporting them immediately to us, at Care of Sweden. 1.2 Intended use The system consists of a pump and a mattress and are used for prevention and as an aid in the treatment of pressure ulcers/pressure injuries (PU/PI).
This guide provides guidance to personnel and prescribers when choosing a mattress from Care of Sweden. CuroCell S.A.M® is a mattress in function group B. This information can be found on our website www.careofsweden.com or you can order a copy from our Customer services by sending your details to export@careofsweden.se.
Page 6
Back side 1. Bed hooks 2. Main power switch (side) For switching on/off the pump. Green lamp indicates when the pump is running. 3. Power cable 2.2 Front panel Comfort settings The comfort setting adjusts the air pressure in the mattress. When the knob is turned clockwise, the air supply and pressure in the mattress increase.
When unpacking the mattress, check that none of the parts are damaged. In the event of damage, contact Care of Sweden or your local distributor before using the product. 1a. If the mattress is an overlay mattress, place the mattress on top of an existing mattress. If possible, secure it by pulling the elastic bands on the bottom over the corners of the underlying mattress.
Page 8
3.3 Comfort settings The air pressure in the mattress can be adjusted according to the user´s desire for comfort. Always consult the prescriber or other authorised personnel for the appropriate setting. Adjust the pressure with the comfort setting knob according to the table on the pump.
Before using the product again for a new user, it is recommended to clean the mattress in accordance with the instructions below. Consult your hygiene manager or Care of Sweden for help and instructions if you are unsure. Cleaning and reconditioning should otherwise be performed as required.
Page 10
INNER COVER Clean the affected area with a mild cleaning solution for this purpose such as: alcohol with or without tenside surfactants, Isopropanol 70 %, oxidizing solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %. 4.2 Reconditioning PUMP See section 4.1 for instructions.
3. Check that air is coming from the outlets on the side of the pump. 4. Check for damaged or pinched tubes and connectors. Contact Care of Sweden or your local distributor for spare parts. 7. Troubleshooting Problem...
80/85/90/105/120x200/210x10/16 cm Weight 2.7 kg NOTE! Care of Sweden reserves the right to modify the product specification at any time. WARNING! Follow the operation and maintenance instructions carefully in order to prevent physical injury, as well as damage to the product or other equipment.
Page 13
8.4 Symbol Definitions XXXX-XX-XX Item no Manufacturer Year-Month-Day Foot placement Read the instructions Group of function CE marked acc. to Medical Recycling for use Device Regulation MDR (2017/745) Recommended user weight User information Not in domestic refuse, Directly on bed base –...
120x210x10 cm Art.No Description CC-250101-E3 CuroCell S.A.M.®, pump without quick connection CC-250102 CuroCell S.A.M.®, pump incl. quick connection CC-908032 Tubing set incl, CPR If other spare parts are needed, please contact Care of Sweden or your local distributor for further assistance.
Pump bag, black 10.3 Other accessories To ensure the best level of comfort, we recommend use of accessories from Care of Sweden since our products are designed to work well together. Please contact Care of Sweden or your local distributor for accessories or more information on our products.
The scope of the servicing and the servicing intervals will depend on how often and to what degree the product is used. Service and maintenance must always be performed by Care of Sweden or one of its authorised technicians. Only use spare parts that have been approved by Care of Sweden. Use of non-approved spare parts invalidates the warranty.
For more information contact Care of Sweden or your local distributor. 12.2 Returns and complaints Complaints are to be reported to the distributor. Return of a product shall follow the distributor’s instructions and can only be done after the complaint has been reported.