CEM Razor Installation And Operation Instructions Manual

CEM Razor Installation And Operation Instructions Manual

Parallel cleavage system
Hide thumbs Also See for Razor:

Advertisement

Quick Links

Razor
Parallel Cleavage System
Installation and Operation Instructions
600850
Rev. 2
July 2017

Advertisement

Table of Contents
loading

Summary of Contents for CEM Razor

  • Page 1 Razor ™ Parallel Cleavage System Installation and Operation Instructions 600850 Rev. 2 July 2017...
  • Page 2 Razor Hardware Vessel Hot block Valve Panel Vessel Tray Temperature Control Panel Locking Lever Power Switches Vacuum Switch Heat Vacuum Switch Gauge Vent, Vacuum, and Waste Ports Razor™ Parallel Peptide Cleavage System 600850 • Revision 2 • July 2017...
  • Page 3: Hardware Installation

    • can accomodate the Razor system [15” x 11.5” x 15.5” (38.1 cm x 29.2 cm x 39.4 cm) (L x W x H)] as well as an additional 8” (20 cm) of open space on each side and 6” (15 cm) at the rear of the instrument for proper ventilation.
  • Page 4 The Razor should be installed in a ventilated fume hood. Remove the Razor from it’s shipping box and place the system in a well ventilated fume hood. Connect the power cord to the power supply box. Attach the power supply to the connector on the rear of the system, and plug the power cord into an appropriate outlet.
  • Page 5: Recommended Cleavage Parameters

    Load the vessel tray with 12 clean 50 mL centrifuge tubes (P/N: 330090), and slide into the Razor cavity. Make sure the tray has been pushed all the way to the back of the cavity. Lock the vessel tray in place by pulling the locking lever down.
  • Page 6: General Procedures

    Verify the current system temperature has reached the set temperature. Load the vessel tray with 12 clean 50 mL centrifuge tubes (P/N: 330090), and slide into the Razor cavity. Make sure the tray has been pushed all the way to the back of the cavity. Lock the vessel tray in place by pulling the locking lever down.
  • Page 7 Move the reaction tube containing the rinsed resin to an empty, clean position in the Razor hot block and turn the numbered valve knob to the CLOSED position. Add cleavage cocktail to the reaction vessel, double that of the volume of the resin bed. For 0.25 mmol cleavages, the resin mixture may be visible above the hot block.
  • Page 8: Maintenance Procedures

    1 L glass waste bottle (no cap) 167545 GL45 waste cap 547225 Waste Tubing Bundle NP-1392 Cleavage vessels, pack of 120 170810 Cleavage vessel cap 170875 Vessel rack cover Razor™ Parallel Peptide Cleavage System 600850 • Revision 2 • July 2017...
  • Page 9: Limited Warranty

    Limited Warranty What Is Covered: CEM Corporation warrants that the instrument will be free of any defect in parts or workmanship and will, at its option, replace or repair any defective part (excluding consumables) or instrument. For How Long: This warranty remains in effect for 365 days from date of delivery to the original purchaser.
  • Page 10 CEM Corporation. CEM® is a registered trademark of CEM Corporation. Razor™ is a trademark of CEM Corporation. www.cem.com • peptide.support@cem.com • (800) 726-3331 (US + Canada) • (704) 821-7015...