CEM DT-982 User Manual

CEM DT-982 User Manual

Thermal imager 160 x 120 pixels auto fusion function

Advertisement

Quick Links

160 x 120
Pixels
AUF: Auto Fusion function
(Patent Num: PCT/CN2016/079015)
Thermal Imager
User Manual
Please read this manual before switching the unit on.
Important safety information inside.

Advertisement

Table of Contents
loading

Summary of Contents for CEM DT-982

  • Page 1 160 x 120 Pixels AUF: Auto Fusion function (Patent Num: PCT/CN2016/079015) Thermal Imager User Manual Please read this manual before switching the unit on. Important safety information inside.
  • Page 2 Thermal Imager User Manual...
  • Page 3: Table Of Contents

    Thermal Imager User Manual Contents Page ............... Introduction ............. Safety Information .............. Packing lists ............Specifications ..........Structure Description ............5.1- Back View ............5.2- Front View ............. 5.3- Interface ............Before You Start ........6.1-How to Charge the Battery ..........
  • Page 4: 2- Safety Information

    Thermal Imager User Manual 1.Introduction overview The Thermal Imager is handheld imaging camera used for predictive maintenance, equipment troubleshooting, and verification. Thermal and visual images are displayed on the LCD and can be saved to a Micro SD Memory card. Transferring images to a PC is accomplished by removing the SD memory card and connecting it to a PC through the included card reader.
  • Page 5: 3- Packing Lists

    Thermal Imager User Manual can also decrease the performance or the life cycle of the battery. Do not get water or salt water on the battery, or permit the battery to get wet. Clean the case with a damp cloth and a weak soap solution. Do not use abrasives, isopropyl alcohol, or solvents to clean the case or lens/screen.
  • Page 6: 4- Specifications

    Thermal Imager User Manual 4.Specifications Imaging and Optical Data Field of view (FOV) / Minimum focus distance 19°x 25°/ 0.5m Spatial resolution (IFOV) 2.78mrad Thermal sensitivity/NETD < 0.1°C @ 30°C (86°F) / 100 mK Image frequency 50Hz Focus mode Manual Zoom 1-32 x continuous, digital zoom Focal length...
  • Page 7: Back View

    Thermal Imager User Manual Data Communication Interfaces Interfaces USB-mini, audio, HDMI Data transform between camera and PC Live video between camera and PC Video out HDMI Power System Battery Li-ion battery, 4 hours operating time Input voltage DC 5V Charging system In camera (AC adapter) Power management Automatic shutdown...
  • Page 8: Front View

    Thermal Imager User Manual 5.2-Front View LED Light Visual Camera Laser Pointer Infrared Camera Lens Hole for Tripod Insertion 5.3-Interface Charge Light USB Cable Connection /Charger Input Terminal Audio/Microphone HDMI Output Micro SD Card...
  • Page 9: 6- Before You Start 6.1-How To Charge The Battery

    Thermal Imager User Manual 6.Before You Start 6.1-How to Charge the Battery Before you use the Thermal Imager for the first time, charge the battery for a minimum of one and one-half hours. The battery status shows on the four-segment charge indicator. To charge the battery, use follow before: Connect the ac power adapter into an ac wall outlet and connect the dc output to the Thermal Imager’s ac power socket, the charge light is on.
  • Page 10: Lens

    Thermal Imager User Manual 6.4-Lens The Thermal Imager has a Lens. is the largest area that your imager can see at a set distance. This table lists the horizontal FOV, vertical FOV and IFOV for lens. Focal Length Horizontal FOV Vertical FOV IFOV 19°...
  • Page 11: Shutter

    Thermal Imager User Manual 6.5-Focus To adjust focus, clockwise or Anti-clockwise rotates the IR Lens. When target comes into focus, it shows a sharper image. When the target moves out of focus, the thermal image becomes blurry. Note Correct focus is important in all imaging applications. Correct focus makes sure that the infrared energy is correctly directed onto the pixels of the detector.
  • Page 12: Temperaturemeasurement

    Thermal Imager User Manual 6.10-Temperature Measurement All objects radiate infrared energy. The quantity of energy radiated is base on the actual surface temperature and the surface emissivity of the object. The Thermal Imager senses the infrared energy from the surface of the object and uses this data to calculate an estimated temperature value.
  • Page 13: Reflected Temperature

    Thermal Imager User Manual 6.12-Reflected Temperature Using the offset factor, the reflection is calculated out due to the low emissivity and the accuracy of the temperature measurement with infrared instruments is improved. In most cases, the reflected temperature is identical to the ambient air temperature. Only when objects with strong emissions with much higher temperature are in the proximity of the object being measured should be determined and used.
  • Page 14: 7-Menus

    Thermal Imager User Manual 7.Menus The menus, together with buttons, are access for image, measurement, Emiss, Palette, temperature measurement range, take photo and video, review, and settings. 7.1-Main Menu Main Menu is the main interface of the Thermal Imager’s menus. It contains six items such as Measure, Emiss, Image, Palette, Temperature measurement range, Settings.
  • Page 15: Image Palette

    Thermal Imager User Manual 7.3-Image Palette The Image Palette lets you change the false-color presentation of the infrared images on display or captured. A variety of palettes are available for specific applications. The standard palettes offer an equal, linear presentation of colors that allow for best presentation of detail. Standard Palette In main menu, press “up”...
  • Page 16: Measurement Menu

    Thermal Imager User Manual 7.5-Measurement Menu In main menu, highlight “Measure” and press “select” button, popup Measure submenu which contains Spot, Hot, Cold. Spot: Measure the center point temperature. Hot: capture maximum temperature. Cold: capture minimum temperature. Open and Close Press “Select”...
  • Page 17 Thermal Imager User Manual Emissivity In Object parameter submenu, press “up” and “down” button, highlight “Emiss”, press “select” button, popup emissivity submenu. “Emiss” sets object emissivity, the value range is 0.01~1.00; Reflective Temperature The reflective temperature is important for radiometric temperature measurement. Thermal Imager has temperature compensation for reflective temperature.
  • Page 18: Temperature Ranges

    Thermal Imager User Manual 7.7-Temperature Ranges The temperature measurement ranges have “-20~150°C” and “0~350°C” to choose. The overlap temperature of the two ranges is more accurate to choose “-20~150°C”. In main menu, press “up” and “down” button, highlight “Range”. Press “select” button, popup temperature ranges submenu. Press “up”...
  • Page 19 Thermal Imager User Manual 7.8.1-Language In Settings submenu, press “up” and “down” button, highlight “Language”. Press “select” button, popup language submenu. Press “up”, “down”, “left” and “right” button, highlight the language which you want to choose. Press “Select” button to confirm, “Close” to exit without change. 7.8.2-Temperature Unit In Settings submenu, press “up”...
  • Page 20 Thermal Imager User Manual 7.8.3-Factory Set In Settings submenu, press “up” and “down” button, highlight “Factory Set”. Press “select” button, popup “Factory Set” submenu. Press “OK” to restore factory setting, “Cancel” to exit without change. Factory Settings of the Thermal Imager is as follow: Item Parameter Value...
  • Page 21 Thermal Imager User Manual 7.8.4-Set Time In Settings submenu, press “Up” and “Down” button, highlight “Set time”. Press “select” button, popup “Set time” submenu. Press “Left” and “Right” button, highlight the item which you want to choose. Press “Up” and “Down” button, modify the value of the selected item. 7.8.5-Information In Settings submenu, press “Up”...
  • Page 22 Thermal Imager User Manual 7.8.6-Alignment In Settings submenu, press “Up” and “Down” button, highlight “Alignment”. Press “select” button, popup “Alignment” submenu. Press “up” and “down” button, highlight the item which you want to choose. Press “Select” to confirm, “Close” to exit without change. 7.8.7-Power off In Settings submenu, press “Up”...
  • Page 23: Camera Menu

    Thermal Imager User Manual 7.9-Camera Menu Thermal Imager has photo and video functions. In photo function, the Imager can save thousands of images. Every image resolution is 1280x960, format is .jpg, and stores infrared data and visible data in an image. In video function, the Imager has .mp4 video capture for hours, and save infrared data in .mp4 format.
  • Page 24: Files Browser

    Thermal Imager User Manual 7.11-Files Browser In desktop, press “Files Browse” button, popup files Browser, Which displays images and videos saved in SD Memory Card. 7.11.1-Open an Image Press “up”, “down”, “left” and “right” button, highlight an image which you want to choose. Press “Open”...
  • Page 25: Usb Mode

    Thermal Imager User Manual 7.11.3-Delete a File or all Files Press “up”, “down”, “left” and “right” button, highlight an image and a video which you want to choose. Press “Option” key, popup “Option” submenu. • Press “Up” and “Down” button, highlight “Delete”, •...
  • Page 26: Hdmi Output

    Thermal Imager User Manual 7.12.2-PC Camera The device is a USB camera for your computer. If select this mode, will display a picture as follow: 7.13-HDMI Output The video output available in the Thermal Imager enables displaying the thermal image(not includes operator menu) on an external monitor or video recording device capable of managing HDMI systems.
  • Page 27: 8-Fault Diagnosis And Exclusion

    Thermal Imager User Manual 8-Fault Diagnosis and Exclusion If you encounter any problems while using the thermal imager, overhaul according to the following table. If the problem persists, disconnect the power and contact with the company's technical support department. Phenomenon of the fault Cause of the fault Solution Thermal imager cannot start...
  • Page 28: Analyse

    Thermal Imager User Manual 10.2- Analyse Select a IR Picture and click “ ” icon to analyse it. 1 . - 0 3 Image mode Click “ ” icon to select image mode, there are four mode for you to select •...
  • Page 29: Analyse

    Thermal Imager User Manual 10.5- Analyse Click “ ” icon to analyse the IR pictures. There are three analyse tools: • Point analyse: Add a point to the picture, it will display the temperature of the point. • Line analyse: Add a line to the picture, it will display the highest, lowest and average temperature of the line. •...
  • Page 30: Share

    Thermal Imager User Manual 11.2- Share Click “ ” icon to share the Infrared picture with Email, Cloud or Message and so on. 12-PC Software Software Install and Uninstall 12.1-System required Window XP or higher version of Windows system, please make sure you have installed Net Framework 2.0 or Net Framework 3.5(include 2.0)when you install PCIMeter software.
  • Page 31: Running

    Thermal Imager User Manual Installation Successful after click “Finish” like above. 12.3-Running After ensuring PCIMeter software has been installed, click shortcuts on the desktop or start menu to run the software. 12.4-Uninstall Uninstall PCIMeter in the start menu as follow, then click "Next"...
  • Page 32 Rev. 160523...

Table of Contents