Welch Allyn Propaq CS 242 Directions For Use Manual page 37

Vital signs monitor
Table of Contents

Advertisement

Directions for Use
To enable the Smartcuf filter:
Connect the ECG leads to the patient and perform ECG monitoring during NIBP.
From the Main Menu, press NIBP to display the NIBP Menu (shown on page 34) and
set Smartcuf to ON.
If artifact is so severe while Smartcuf is enabled that it affects the accuracy of an NIBP
measurement, that measurement is marked with a special symbol on the display and on
printouts:
This symbol indicates the NIBP
reading was taken in the presence of
high artifact while monitoring ECG
with Smartcuf enabled. Artifact can
affect accuracy. To help reduce
artifact, see step 6 on page 35.
NIBP
S
D
18:45
(M)
mmHg
Numeric
Display
There may be some situations where it is desirable to disable Smartcuf. This may include
situations with very extreme motion artifact, certain types of arrhythmias, or other
situations where it is not possible to obtain a good ECG signal. NIBP measurements can
still be performed when Smartcuf is disabled.
To disable Smartcuf, from the Main Menu press NIBP to display the NIBP Menu and set
Smartcuf to OFF.
Take the NIBP reading
1. Select a cuff and hose appropriate for the patient. Select cuff size based on limb
circumference.
Typical hoses
Typical cuffs
Recommended limb
circumference
112
58
( 85)
Printout
WARNING Use only accessories approved by Welch Allyn. Visit
www.welchallyn.com.
The use of any other accessories can result in inaccurate
patient data, can damage the equipment, and can void your product warranty.
Neonate Mode
Neonate/Infant
Neonate #1 to #5
(disposable); newborn,
infant (reusable)
up to 15 cm
NIBP TREND
TIME
HR/PR SpO
2
S D (M)
HH:MM
BPM
%
mmHg
122 58 ( 85)
10:01
80
97
9:46
80
97
126 62 ( 86)
9:31
80
97
127 62 ( 86)
97
134 66 ( 89)
9:20
80
124 57 ( 86)
9:08
80
97
127 58 ( 88)
8:52
80
97
8:47
80
97
124 56 ( 85)
8:41
80
97
126 57 ( 84)
127 58 ( 85)
8:35
80
97
132 61 ( 89)
8:30
80
97
PRINT
Trend Display
Pediatric Mode
Adult Mode
Adult
Adult
Newborn, infant, small child,
Child, small adult, adult,
child, small adult
large adult, thigh
7.7 to 25 cm
greater than 15 cm
Monitoring
33
P
2
(M)
( 15)
mm
RR/BR
NIBP
Br/M
(M)
12
( 85)
mm
12
TEMP
12
12
F
12
MCO
2
12
38
12
mmHg
12
S
p
O
2
12
12
%
NEXT
P
TREND
OX

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Propaq cs 244Propaq cs 246

Table of Contents