114
Reference
Power Specifications
Characteristic
Mode of Operation
Battery Pack Type
Battery Pack Capacity
Battery Recharger Circuitry
DC Input Power Required
Input Fuse Rating
Battery Recharge Time with
instrument on
Battery Recharge Time with
instrument off
Recharge time until monitor is
usable, starting with discharged but
non-faulty battery
Low Battery Voltage and Operation
Monitor only
Monitor and SpO
Monitor with Expansion Module with
printer, SpO
Specification
Continuous
Sealed, gel-type lead acid
Monitor only: 8 V, 2.7 Ampere-Hours; Monitor with Expansion Modules: 8 V, 5.4
Ampere-Hours
Internal, powered by external power adapter
12 to 28 V, 25 Watts
3A/250V, Type 2AG (0.57x 0.177 in)
Range of 8 hours to 12 hours typical, depending upon product configuration
Range of 6 hours to 8 hours depending upon product configuration
2 minutes typically (longer time required before NIBP, printer, and CO
available)
< 7.8 V: Caution message LOW BATTERY.
< 7.6 V: Caution messages LOW BATTERY, PRINTER DISABLED and LOW
BATTERY, NIBP DISABLED.
< 7.4 V: Equipment alert VERY LOW BATTERY
< 7.3 V: Equipment alert LOW BATTERY, HEATER DISABLED (MCO
< 7.0 V: Monitor automatically turns off.
Typical operating times on battery
2 hours
(Baqpaq)
5 hours
2
3 hours
and CO
Options
2
2
Welch Allyn Propaq CS Vital Signs Monitor
)
2
are
2