Power Adapter Specifications
Characteristic
Type of Protection Against Electric Shock
Degree of Protection Against Harmful Ingress
of Water
Method of Disinfection
Flammable Anesthetics
Operating Temperature
Shipping and Storage Temperature
Operating Altitude
Shipping and Storage Altitude
Operating Relative Humidity
Shipping, Storage Relative Humidity
Shock
Vibration
Length
Width
Height
Weight
Rated Input
Rated Fuses
Rated Output (Continuous)
Additional Features
Rated Input
Rated Fuses
Rated Output (Continuous)
Additional Features
a.
Per EN 60601-1 unless otherwise stated.
Specification
Protection classifications, all adapters
Class I, (Protectively Earthed)
For ordinary, indoor locations only.
Not suitable for autoclaving
Not suitable for use with flammable anesthetics
Environmental specifications, all adapters
0° to 50° C
-20° to 60° C
-2,000 to 15,000 feet (-610 to 4,572 m)
-2,000 to 40,000 feet (-610 to 12,192 m)
15% to 95%, noncondensing
15% to 95%, noncondensing
50 g
Random Vibration, 0.02 g
2
g
/Hz at 500 Hz. Operating 1 hour per axis, 3 hours/ test.
Physical specifications
5.0 in (12.7 cm)
3.6 in (9.1 cm)
3.1 in (7.9 cm)
3.1 lb (1.4 kg)
Universal power adapter, part No. 503-0054-00
100-120 V ac, 500 mA, 50/60 Hz
T800 mA/250 V, Time-Delay, 5x20mm
16-24 V dc, 25 VA
Detachable power cord, pilot light
Universal power adapter, part No. 503-0054-01
200-240 V ac, 250 mA, 50/60 Hz
T400 mA/250 V, Time-Delay, 5 x 20mm
16-24 V dc, 25 VA
Detachable power cord, pilot light