Welch Allyn Propaq CS 242 Directions For Use Manual page 110

Vital signs monitor
Table of Contents

Advertisement

106
Reference
Characteristic
Sensor Type
Principle of Operation
Warm-up time (CO
Response Time
Waveform Rise Time
Calibration
Sensor Housing Temperature
Sensor Height
Sensor Width
Sensor Depth
Sensor Weight
Cable Length
Type
Size
Material
Added Deadspace
Operating Ambient Temperature
Storage Temperature
Operating Altitude
Storage Altitude
Operating and Storage Humidity
Shock
Vibration
Drop
a.
Not including cable
Mainstream CO
Specification
Mainstream
Non-dispersive, infrared, single-beam, single path/wavelength, ratiometric
sensor and monitor)
45 s typical, 3 min maximum
2
30 ms typical, 60 ms maximum
<120 ms to 90% after step change
Verify semi-annually, calibrate only as required
42°C nominal
Mainstream CO
2
a
1.003 in (2.548 cm)
1
1.036 in (2.631 cm)
1
0.78 in (1.981 cm)
1
< 0.53 oz (15.03 g)
10 ft (3.05 m) nominal
Mainstream CO
Per ISO 3040, single-use
15 mm ID, (meets ISO specifications)
clear polycarbonate, with sapphire windows
< 6cc (0.37 cubic inches) for adult model, <0.6 cc (0.037 cubic inches) for low
deadspace model
Mainstream CO
10° to 40°C
-20° to 60°C
-2,000 to 15,000 ft (-610 to 4,572 m), 817 to 429 mmHg
-2,000 to 40,000 ft (-610 to 12,192 m), 817 to 141 mmHg
0% to 95%, noncondensing
100 g for 4 ms
5-35 Hz, 0.015 in (0.038 cm) peak-to-peak,
35-100 Hz, 1 g acceleration
36 inches free fall to floor (tile over concrete, one drop each face, one drop
each edge/corner)
Welch Allyn Propaq CS Vital Signs Monitor
specifications
2
Mainstream CO
sensor
2
sensor and cable dimensions and weight
airway adapter
2
sensor environmental specifications
2