7
Maintenance
Introduction to Maintenance and Repair

Using the ESD Strap

Electronic boards are sensitive to electronic discharge (ESD). In order to
prevent damage, always use an ESD strap when handling electronic boards
and components.
1 Unwrap the first two folds of the band and wrap the exposed adhesive side
2 Unroll the rest of the band and peel the liner from the copper foil at the
3 Attach the copper foil to a convenient and exposed electrical ground.

Cleaning the Instrument

Liquid dripping into the electronic compartment of your module.
WA R N I N G
Liquid in the module electronics can cause shock hazard and damage the module.
The micro vacuum degasser case should be kept clean. Cleaning should be
done with a soft cloth slightly dampened with water or a solution of water and
a mild detergent. Do not use an excessively damp cloth that liquid can drip
into the micro vacuum degasser.
48
firmly around your wrist.
opposite end.
Figure 12
Using the ESD Strap
Do not use an excessively damp cloth during cleaning.
Drain all solvent lines before opening any fittings.