Theory Of Operation; Esd Protection At The Probe Tip - Agilent Technologies 85024A User's And Service Manual

High frequency probe
Table of Contents

Advertisement

Service

Theory of Operation

Theory of Operation
The probe uses a Gallium Arsenide FET integrated circuit amplifier which provides unity gain. This
amplifier microcircuit requires +6 V and 4 Vdc to operate. These voltages are provided by the regulator
assembly.
The regulator assembly converts the +15 volt and 12.6 volt supplies from the host instrument with two
voltage regulators. The regulator supplies the +6 and 4 voltages to the amplifier microcircuit.
The most common failures in the probe will all result in loss of signal through the probe, and will show
up if the operator's check is performed. The most common failures are expected to be:
1. Destruction of the amplifier microcircuit due to static electricity (proper anti-static precautions not
taken).
2. Probe power tip damage (caused by the operator dropping the probe with the sleeve retracted).
3. Possible regulator failure.

ESD Protection at the Probe Tip

The FET integrated circuit amplifier is located at the tip of the probe. The FET input of the amplifier is
protected by a bidirectional voltage limiter as shown in
Figure
7-1.
Figure 7-1
Probe Tip Input Circuit
The voltage limiter element has a negligible effect on circuit operation for input voltages within
specification. The voltage limiter is a non-linear element: it begins to conduct current at about 2.5 volts.
It effectively limits voltages at the amplifier input to less than 4 volts. This prevents amplifier damage
from low levels of ESD.
Even with the protection provided by the voltage limiter element, the probe is still sensitive to ESD.
High levels of ESD can cause permanent damage to the voltage limiter element itself. Careful ESD
precautions must be observed when using the probe.
Chapter 7
7-3

Advertisement

Table of Contents
loading

Table of Contents