Installing the Micro Vacuum Degasser

Operational Hints for the Micro Vacuum Degasser

Operational Hints for the Micro Vacuum Degasser

General Priming Instructions

When opening capillary or tube fittings solvents may leak out.
WA R N I N G
The handling of toxic and hazardous solvents and reagents can hold health risks.
Before using a new degasser or new tubings for the first time:
1 Prime all tubings with at least 5 ml of iso-propanol no matter whether the
2 Replace the current solvent with iso-propanol, if current solvent is organic
24
Please observe appropriate safety procedures (for example, goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are
used.
channels will be used with organic mobile phase or with water.
If you are changing to a solvent that is immiscible with the solvent currently
in the tubing continue as follows:
or with water, if current solvent is an inorganic buffer or contains salt.
1260 Infinity Micro Degasser User Manual