Leakwise ID-220 SERIES User Manual page 76

Smart leakwise controller & smart leakwise receiver for remote slc-220’s
Table of Contents

Advertisement

10. Turn SLC-220 power off, wait 5 seconds, and then turn power on again. Run the SLC-
Manager PC program and verify that the version (shown right to the Leakwise logo) is
correct.
11. Load from PC the saved settings, and then perform Save to SLC.
12. Verify that the settings are correct in all SLC-Manager screens. Click Change Parameters
in each screen, including Advanced screens, and check if there are parameters in red or
blue color. If yes (except Sensor Name), do a Save to SLC again. Finally, verify that the
system is operating correctly.
Note:
1. The repeating lines stating "modbus error RX_TIMEOUT" are a normal behavior during the initial
communication between the PC and the SLC-220.
The "RX Password:
connection to the Main board.
SLC-220 User Guide Rev 2.0
DONE" line and then "Changing baud rate..." indicate a successful
76
June 2018