1. In each screen where parameter where changed, press Save to SLC button or Save to
PC button (as appropriate) BEFORE changing to a different screen or exiting the
program. If a substantial sensor parameters change is done in the two screens related to
the sensor, press the Save to SLC button only after doing all such changes, to avoid
conflicts between parameters.
2. On versions before 2.0.0.1: After Load From PC, do a Save to SLC from each screen.
3. It is recommended to have the parameters saved also to the PC for future reference.
4. When a parameter is changed, it is colored red until it is saved to the SLC or to the PC.
5. If the Save operation is stuck or gives an error message, reset the SLC-220 and the SLC-
Manager.
6. After the parameters are saved on the SLC-220 controller, the on-board Buzzer will
confirm it with a single beep, or reject the data by sounding multiple beeps.
Pressing the Advanced button enables an Administrator user to view and change additional
special parameters:
Pressing the View Parameters button returns the display to the basic Sensor parameters.
1. Alarm Reading Percentage Threshold: Relevant for firmware versions before 4.1. When
the Delay Time (Filter) opens a measurement window to filter out unstable readings, the
Alarm Reading Percentage Threshold (between 5% and 100%) defines the proportion
between the two alternating statuses for determining the "winning" status in unstable liquid
conditions. For example, if 70% is selected, and then 85% of the readings "vote" Oil status
while the other 15% "vote" Water status, then Oil status will be determined at the end of
the Delay Time.