Specifications
Table 4: Trigger specifications (cont.)
USB
Ethernet
Digital Acquisition Specifications, MSO4000B Series
Table 5: Digital acquisition specifications, MSO4000B Series
Characteristic
Threshold voltage range
Digital channel timing resolution
Logic threshold accuracy
Minimum detectable pulse width,
typical
18
Data Rates Supported: HS: 480 Mbps, Full: 12 Mbps, Low: 1.5 Mbps
Trigger On: Sync, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data
Packet, Handshake Packet, Special Packet, Error
NOTE.
HIGH SPEED support available only on MSO4104B, DPO4104B, MSO4104B-L,
DPO4104B-L, MSO4102B-L, DPO4102B-L, MSO4102B, and DPO4102B models.
Bit Rate: 10BASE-T, 10 Mbps; 100BASE-TX, 100 Mbps
Trigger On: Start Frame Delimiter (SFD), MAC Address, MAC Length/Type, IP Header, TCP
Header, TCP/IPv4/MAC Client Data, End of Packet, Idle, FCS (CRC) Error, MAC Q-Tag
control Information.
The following table shows the digital acquisition specifications for the MSO4000B
Series oscilloscopes.
Description
-40 V to +40 V
2 ns main memory, 60.6 ps for MagniVu memory
±(100 mV + 3% of threshold setting after calibration)
Requires valid SPC, as described in step 2 of the Self Test. (See page 55, Self Test.)
1 ns
Using MagniVu memory. Requires the use of 342-1140-00 ground clip on each channel.