Rosemount 6888Xi Quick Start Wizard; Re-Initiating Rosemount 6888Xi Wizard - Emerson Rosemount 6888C Instruction Manual

In-situ combustion oxygen analyzer for use in hazardous areas
Hide thumbs Also See for Rosemount 6888C:
Table of Contents

Advertisement

Instruction Manual
00809-0100-4891
5.4

Rosemount 6888Xi Quick Start Wizard

When the Rosemount 6888Xi is first powered, a short wizard program guides you through
the basic setup. Once configured, the Rosemount 6888Xi retains the setup, and the wizard
will not repeat.
1. Apply power to the Rosemount 6888Xi.
Once boot-up is complete, the Quick Start Wizard screen appears. With a dual
channel Rosemount 6888Xi, the wizard runs for both channels in succession.
2. Press Enter to continue.
3. At the Sensor Type screen, use the Up and Down keys to select O
Do not select CO as this option is reserved for future use.
4. Press Enter to continue.
5. At the Device Type screen use the Up and Down keys to select HART
(FOUNDATION
6. At the Auto Cal Device screen, use the Up and Down keys to select the calibration
method to be used. The methods are defined as follows:
7. Press Enter to continue.
NOTICE
If SPS, IMPS, or Integral is selected, you must still configure automatic calibration as
On. Other parameters, such as test gas values and gas times, should be verified as well.
Refer to the
8. When prompted by Setup Correct? , use the Up and Down keys to select Yes.
If you select No, the wizard restarts.
9. Press Enter to continue.
The Rosemount 6888Xi displays several screens while saving the configuration, resets
itself, and then returns to the main screen.
5.5

Re-initiating Rosemount 6888Xi wizard

To re-initiate the wizard setup, you must reset the I/O board to default factory conditions.
1. Apply power to the Rosemount 6888Xi.
2. When the main screen appears, press Menu several times until the System menu
appears. Use the Up and Down keys to select Configure IOB. Press Enter to continue.
3. When the Configure IOB screen appears, use the Up and Down keys to select I/O
Board 1. With a dual channel Rosemount 6888Xi, you can reset and reconfigure either
or both I/O boards. Press Enter to continue.
Instruction Manual
Fieldbus), whichever applies.
None - Manual calibration with the standard probe housing configuration
SPS - Automatic calibration with the standard probe housing configuration using
the Rosemount SPS 4001B
IMPS - Automatic calibration with the standard probe housing configuration using
the Rosemount IMPS
Integral - Automatic calibration with the integral autocal probe housing
configuration
Rosemount 6888Xi Reference Manual
.
2
®
or FF
for calibration setup details.
Power up
July 2018