Maintenance and service
July 2018
8.3.6
Diffusion element replacement
The diffusion element protects the cell from particles in process gases. Normally, it does
not need to be replaced, because the vee deflector protects it from particulate erosion.
In severe environments, the diffusion element may be broken or subject to excessive
erosion. Examine the element whenever removing the probe for any purpose. Replace if
damaged.
Damage to the diffusion element may become apparent during calibration. Compare
probe response with previous response. A broken diffusion element will cause an
inadequate response to calibration gas. Hex wrenches needed to remove setscrews and
socket head screws in the following procedure are available as part of a probe disassembly
kit,
Table 9-1
NOTICE
This refers to the ceramic diffusion element only.
74
on page 79.
Instruction Manual
00809-0100-4891
6888C