Replacement screws are provided if any become lost or damaged.
9. Reconnect the power and signal wires to the transmitter board.
10.Reinstall the cover to the DR probe housing.
11.Reapply power to the DR probe.
8.3.4
Heater strut replacement
This section covers heater strut replacement. Do not attempt to replace the heater strut
until all other possibilities for poor performance have been considered.
If heater strut replacement is needed, order a replacement heater strut kit
page 79). Refer to
WARNING
BURNS
Use heat resistant gloves and clothing when removing probe. Do not attempt to work on
the probe until it has cooled to room temperature. The probe can be as hot as 427 °C (800
°F). This can cause severe burns.
Instruction Manual
Figure 8-3
on page 70 for item letter references.
Maintenance and service
July 2018
(Table 9-1
on
69