00809-0100-4891
Figure 4-3: Single/dual channel wiring diagram
NOTICE
1. Except for JP5, JP7, and JP8 on IO board, jumper and switch settings are factory set and are shown for reference only.
2. IO board 4-20 mA/HART
JP5
Pins 1-2: internal loop power Rosemount 6888 Xi to Rosemount 6888 Transmitter
Pins 2-3: external power Rosemount 6888 Xi to Rosemount 6888 Transmitter (requires 2,500 resistor across
J4 PR+ to PR-)
JP7/JP8
Pins 1-2: internal power from Rosemount 6888 Xi to DCS
Pins 2-3: external power Rosemount 6888 Xi to DCS
Instruction Manual
®
loop power settings