Device Status:
Rosemount 5300 Series
Table 5-1. Device status.
Running Boot Software
Device Warning
Device Error
Simulation Mode 0
Simulation Mode 1
Invalid Measurement
User Register Area Write
Protected
Write Protected Jumper Set
Factory Settings Used
Probe missing
Rosemount Radar Level Transmitters
Device Status messages that may appear on the Integral Display, on the 375
Field Communicator, or in the Rosemount Radar Master (RRM) program are
shown in Table 5-1:
Message
The application software could not be
started.
A device warning is active.
A device error is active.
The simulation mode is active.
The advanced simulation mode is
active.
The level measurement is invalid.
The configuration registers are write
protected.
Write protection jumper on the display
is enabled.
The factory default configuration is
used.
Probe is not detected.
Description
Contact Emerson Process
Management Service Department.
See Warning Messages for details.
See Error Messages for details.
Turn off the simulation mode.
To turn off the Advanced Simulation
mode, set Holding Register 3600=0.
Check Error Messages, Warning
Messages and Measurement Status
for details.
Use the Lock / Unlock function to turn
off the write protection.
Remove the write protection jumper.
The transmitter has lost its calibration.
Contact Emerson Process
Management Service Department.
Check that the probe is correctly
mounted. Check the connection
between probe and transmitter head.
Action