13. Replacing The Filter & Sensor
The sensor needs to be replaced when it is unable to be calibrated or behaves erratically. Also, the
filter should be inspected periodically, and it should be replaced if it shows signs of accumulated
dust, dirt, or debris.

13.1. Filter Color Reference

One of the holes in the sensor cover has a clear plastic window. While the other holes allow
ambient air (and dirt, debris, etc.) to reach the filter, the window keeps the part of the filter under it
clean. As a result, it is easy to visually confirm that the filter needs replacement by simply
comparing the clean spot with the other spots.

13.2. Removing The Cover

Remove the front cover by using the tab end of the calibration cap or a flat-blade screwdriver to pry
it upwards from the slot. This exposes the sensor with its filter attached.
Align tab
Insert tab
With slot
Honeywell BW Solo User Manual
IMPORTANT!
Reference
Window
Clean Filter
Press
Down
In slot
Dirty Filter
Pop up
Cover
36
Reference
Window
Sensor
with filter