For example.: "9256 BWS 0A 18 30 00001" indicates that the instrument was manufactured in the
30th week of the year 2018.
17. Sensors & Settings
The available sensors and their range, resolution, temperature range, and low and high setpoints
are shown here:
Gas
Range
Type
H
S*
0-200 ppm
2
CO*
0-2000 ppm
O
*
0-30% v/v
2
H
S
0-500 ppm
2
(extended
range)
CO-H
0-2000 ppm
NH
0-100
3
NH
0-400
3
(extended
range)
SO
0-100 ppm
2
HCN
0-100 ppm
Cl
0-50 ppm
2
NO
0-250 ppm
NO
0-100 ppm
2
PH
0-5 ppm
3
ETO
0-100 ppm
ClO
0-1 ppm
2
O
0-1 ppm
3
H
0-1000 ppm
2
* 1-Series sensor
Honeywell BW Solo User Manual
Resolution
0.1
1 ppm
0.1 %v/v
0.1*
0.5 ppm
1 ppm
1 ppm
0.1 ppm
0.1 ppm
0.1 ppm
0.2 ppm
0.1 ppm
0.1 ppm
0.1 ppm
0.01 ppm
0.01 ppm
2 ppm
Temperature
-40 to +60° C/-40 to 140°F
-40 to +60° C/-40 to 140°F
-40 to +60° C/-40 to 140°F
-40 to +50° C/-40 to 122°F
-30 to +50° C/-22 to 122°F
-20 to +40° C/-4 to 104°F
-20 to +40° C/-4 to 104°F
-30 to +50° C/-22 to 122°F
-20 to +50° C/-4 to 122°F
-20 to +40° C/-4 to 104°F
-30 to +50° C/-22 to 122°F
-20 to +50° C/-4 to 122°F
-20 to +50° C/-4 to 122°F
-30 to +50° C/-22 to 122°F
-20 to +40° C/-4 to 104°F
-30 to +50° C/-22 to 122°F
-20 to +50° C/-4 to 122°F
44
Low
High
Setpoint
Setpoint
10 ppm
15 ppm
35 ppm
200 ppm
19.5 %v/v
23.5 %v/v
10 ppm
15 ppm
35 ppm
200 ppm
25 ppm
50 ppm
25 ppm
50 ppm
2 ppm
5 ppm
4.7 ppm
10 ppm
0.5 ppm
1.0 ppm
25 ppm
25 ppm
2 ppm
5 ppm
0.3ppm
1.0 ppm
1 ppm
5 ppm
0.10 ppm
0.30 ppm
0.10 ppm
0.20 ppm
100 ppm
500 ppm