Integrator Calibration

Refer to the Integrator instruction manual for Integrator Calibration instructions.

Zero Calibration

Refer to the Integrator instruction manual for Zero Calibration instructions.

Span Calibration

After a successful zero calibration, apply the test weight for the span calibration:
1.
Attach one end of a string (monofilament fishing line or fine flexible cable) to the
test weight.
2.
Route the other end of the string over the calibration pulley.
3.
Attach the free end of the string to the string fixing bolt. Ensure the string rests in the
groove of the bolt.
Note:
Ensure there is no material flow during zero and span calibrations
Ensure the test weight is suspended free of any obstructions.
Accurate calibration is not assured until material tests and a manual span
adjustment have been performed, as outlined in the Integrator instruction manual.