Troubleshooting - Siemens milltronics ILE-37 Instruction Manual

Sensing head
Hide thumbs Also See for milltronics ILE-37:
Table of Contents

Advertisement

Troubleshooting

Every Milltronics ILE-37 sensing head is subjected to extensive quality assurance
procedures to ensure the highest possible degree of quality, reliability, and performance
is achieved.
The following listing indicates the probable cause, and proper course of action to be
taken should the specified fault symptom occur.
Symptom
Integrator rate display doesn't
change when sensing plate is
moved.
Span adjustment does not have
enough range.
Measurement results are not
repeatable.
Accuracy varies with material
flowrate.
Page 14
Wrong or bad integrator
Viscous damper cover in
shipping position.
Integrator not prepared for
Range spring not suited to
Sensing head not level.
Moving beam travel is
mechanically limited.
Leaf springs are damaged.
Material flow patterns vary.
Non-linear operation
Milltronics ILE-37 – INSTRUCTION MANUAL
Cause
connection.
operation.
application
Action
Refer to INSTALLATION/
Interconnection
Refer to INSTALLATION/
Viscous Damper.
Program and calibrate the
integrator.
Refer to RANGE SPRINGS.
Refer to INSTALLATION and
CALIBRATION/Sensing Head
Level Test
Ensure moving beam does not
hit travel stops between –20%
and 150% flowrates.
Replace leaf springs, re-
calibrate Flowmeter and
Integrator.
Refer to the flowmeter
instruction manual
APPLICATIONS section.
Refer to LINEARITY.