Grounding

The equipment must be grounded to reduce the
risk of static sparking and electric shock. Electric or
static sparking can cause fumes to ignite or
explode. Improper grounding can cause electric
shock. Grounding provides an escape wire for the
electric current.
1. Ground the flow meter by connecting a grounded
cable to the meter housing or sensor.
Have a qualified electrician check the electrical
grounding continuity between the flow meter sensor
and a true earth ground; remove the cable connec-
tor from the sensor and measure the resistance
from the cable connector Pin 2 to true earth ground.
Refer to F
. 3.
IG
If the resistance is greater than 25 ohms, check the
cable ground connection. Refer to F
the ground sheath or replace the cable. Do not oper-
ate the system until the problem is corrected.
2. Always ground the meter, using one of the following
options:
Cable Part No.
17C743
17C910
17C887
17C888
17C889
24Y434
NOTE: These cables have a metal connector, which will only mate with a sensor that
also has a metal connector. Adaptor 24Y434 is required to connect an existing cable
with a plastic connector to a replacement sensor, which has a metal connector.
Adaptor 24Y434 purchased separately.
308778Y
. 3. Reconnect
IG
Table 1: Meter Cables
Length
5 ft (1.52 m)
42 in. (1029 mm)
6 ft (1.83 m)
50 ft (15.25 m)
150 ft (45.75 m)
Adaptor - Not included with cables (see note)
a. Mount the meter to a grounded conductive sur-
face, or
b. Connect the conductive fluid hose to the meter
inlet and outlet, or
c. Connect a ground wire to the meter's M6
mounting holes.
3. Never use the flow meter with an electrostatic gun
isolation stand.
3
1
2
F
. 3
IG

Meter Cables and Adaptor

See Table 1 for available meter cables or the adaptor for
use with cables with a black plastic connector and a
replacement sensor.
Ferrite
Platform
Various
15D906
ProMix 2KS
Various
Various
Various
Installation
1
+10-30 Vdc Supply (red)
2
Ground (black)
3
Signal Out (white)
TI26838a