Relay Output - Dräger X-zone 5000 Instructions For Use Manual

Hide thumbs Also See for X-zone 5000:
Table of Contents

Advertisement

RS485 connection:
1
2
XEXT2
3
1
1
2
2
XEXT1
3
3
CAUTION
Ensure that the cable strands are not crossed!
1
1 Plug (male):
Housing: Souriau UTS6JC147P (male)
Contacts: Souriau RM20M12K (male)
2 Plug (female):
Housing: Souriau UTS6JC147S (female)
Contacts: Souriau RC20M12K (female)
3 Cable type: Belden 3107A, 2 x 2 AWG 22
Cable length: max. 25 m between 2 devices
1
2
3
1
2
3
3
max. 25 m
2
DRAFT 02 - 08/02/10

Relay output:

5
1
NO
2
3
6
NC
4
The relay output is intended for connection to an
isolation amplifier with intrinsically safe output.
It is essential that the parameters of the relay outputs
are observed to prevent the circuit from jeopardising
the intrinsic safety of the device.
Only intrinsically safe circuits are permitted for
connection.
The person responsible for the use of the Dräger X-
zone 5000 must prepare a system document to verify
the intrinsic safety level.
2
1 Plug (female):
Housing: Souriau UTS6JC147S (female)
Contacts: Souriau RC20M12K (female)
2 Configure the pin assignment separately depending
on the application
3 Cable type: Belden 3107A, 2 x 2 AWG 22
Cable length: max. 25 m between 2 devices
CO
WARNING
NOTICE
6
XEXT2
5
4
1
2
3
3
max. 25 m
1
11