Set High Frequency Detection averages
For analysis parameters with HFD or VHFD defined as the parameter type, Set HFD Avgs
lets you set the number of averages the analyzer uses when making these measurements.
A value of 16 or larger is recommended. Averaging helps reduce noise.
Procedure
1.
2.
3.
4.
4.3.4
Set the route data storage mode
Data Storage Mode sets how the analyzer saves collected data. You can set the analyzer to
save all data or to overwrite the previous data.
Procedure
1.
2.
3.
MHM-97432 Rev 7
Activate a route.
From the Route Data Collection screen, press ALT > F1 User Setup > F3 Point Advance
Mode.
Enter the number of seconds between 0 and 30.
To disable the feature, enter 0. The default is 0.
Press Enter.
Activate a route.
From the Route Data Collection screen, press ALT > F1 User Setup > F4 Set HFD Avgs.
Enter a number between 1 and 99.
The default is 25.
Press Enter.
Activate a route.
From the Route Data Collection screen, press ALT > F1 User Setup > F5 Data Storage
Mode.
Press F5 Data Storage Mode to scroll through the options.
Option
Description
Always Overwrite Save one set of route data per measurement point, and overwrite any old
data. The default is Always Overwrite.
Query Overwrite
Save only one set of route data per point, and ask before overwriting data.
Save new route data and keep old data stored on a measurement point.
Save All Data
You can keep multiple sets of data collected for a point.