Appendix 2; Turbidity Values Under 1 Fnu/Ntu - wtw Turb 430 IR/T Operating Manual

Handheld turbidimeter
Table of Contents

Advertisement

Calibration procedure:
Measuring
70

Appendix 2:

Turbidity values under 1 FNU/NTU

With turbidity values (under 1 FNU/NTU), the measured value is
strongly influenced by the cell and its alignment.
In order to increase measuring accuracy with turbidity values under 1
FNU/NTU, calibration in the 0.02 FNU/NTU standard and later
measurement should be carried out in the same cell. For calibration in
the standards 10.0 and 1000 FNU/NTU follow the instructions on the
display.
Proceed as follows to measure turbidity values under 1 FU/NTU:
1
Press the <CAL/ZERO> key.
The menu-guided calibration begins.
2
Carry out calibration of the standards 1000 FNU/NTU and
10.0 FNU/NTU in a clean and unscratched cell according to
the menu guidance.
3
Fill the cleaned cell with the 0.02 FNU/NTU standard and
calibrate.
4
Mark the alignment of the cell.
Calibrate
 after the calibration interval has expired
 with a temperature change
5
Fill the marked and cleaned cell with test sample, align it
with the marking and measure.
6
If necessary, fill the marked and cleaned cell once more
with test sample and carry out further measurements.
Turb 430 IR/T
ba75507e05
07/2016