Start guided installation
1. Push menu. Select Maintenance>Installation (scroll up). Complete the steps that show on the
display. The display screens are shown for reference in
colorimeter
on page 74.
2. Wait for the analyzer to shut down. When the display is off, disconnect the power cord from the
electrical outlet.
Note: For user safety, the analyzer will automatically go back to idle mode if the power cord is not disconnected
in 5 minutes or less after the display goes off.
Note: When power is removed from the analyzer for more than 10 minutes, a power-loss beep is heard at the
remote indicator for approximately 5 minutes.
3. Remove the pump clamp from the reagent delivery module.
Appendix A: Screens—Install the
English 49