4.0  CALIBRATION 
 
 Press the [Zero] key and power up the scale using the on/off button. 
 Release  the  [Zero]  key  and  the  display  will  show  "SPAN".  Make  sure  the  hook  is 
empty unless you intend to use any chains, slings etc to hold the calibration mass. 
Add any items such as these to the hook before continuing. 
 Press  the  [Zero]  button  and  the  display  will  show  "SET"  then  "00000"  when  the 
reading at zero is stable. This completes the zero calibration. 
 Following  zero  calibration  the  value  of  the  mass  being  used  to  calibrate  the  scale 
should be entered. Use the [Hold] button to increment the right hand digit, and the 
[Unit] button to move this digit to the left. Proceed by entering a number for the next 
digit  if  required,  until  the  value  of  your  chosen  calibration  weight  is  displayed 
correctly.  
 Load this mass onto the hook and when the scale is stable press the [Zero] key. The 
display shows 'HOLD'. 
 After  a  few  seconds,  the  calibration  is  complete  and  the  scale  returns  to  normal 
weighing showing the value of the calibration mass as used. 
 
** To calibrate in Lbs make sure that the scale is Lbs mode before commencing, to calibrate 
in Kg make sure the scale is in Kg mode before commencing. 
 
 