Fingertip Testing
For fingertip sampling, adjust the depth penetration to reduce the discomfort. You
do not need the clear cap for fingertip sampling.
1. Unscrew the lancing device cover from the body of the
lancing device. Insert a sterile lancet into the lancet
holder and push it until the lancet comes to a complete
stop in the lancing device.
2. Hold the lancet firmly in the lancet holder and twist the
safety tab of the lancet until it loosens. Then pull the
safety tab off the lancet. Save the safety tab for lancet
disposal.
3. Carefully screw the cover back onto the lancing device.
Avoid contact with the exposed needle. Make sure the
cover is fully sealed on the lancing device.
4. Adjust the puncture depth by rotating the lancing device
cover. There are a total of 6 puncture depth settings. To
reduce discomfort, use the lowest setting that still
produces an adequate drop of blood.
22