Bindicator SonoTracker Ultrasonic Liquid Level Installation & Operation Manual page 89

Can monitor up to a total of 8 transducers of various frequencies, accommodating multiple vessels of different heights and shapes.
Table of Contents

Advertisement

www.bindicator.com
4. UNITS OVERRANGE
Explanation: The calculated level exceeds the maximum number for a 16-bit system (65535).
Solution a. If you selected special display units (Spec Display in the Display Menu) for the channel, check the
Format and Maximum Value. A Format of xxx.xx yields a maximum value of 65.535, while a Format of xxxxx00
yields a maximum value of 6553500. Make sure that the Format and Maximum Value is consistent with the
maximum expected value.
Solution b. Check the calibration for the channel, fullpoint and span.
5. MATH ERROR # OR SCRIPT COMPILE ERROR #
Explanation: These messages apply to math channels only. They indicate a problem with the input equation.
Solution: See appendix F, for a detailed list of the math error messages and solutions.
6. NVRAM FAILURE
Explanation: This error message will only occur at power-up. NVRAM data has been corrupted.
Solution: Contact Bindicator® Technical Support at 1-800-426-9010 or 864-574-2763 or default the unit.
7. NOISE ^
Explanation: The amount of noise exceeds noise limit.
85