Hanna Instruments HI 8014 Instruction Manual

Hanna Instruments HI 8014 Instruction Manual

Portable ph meters
Hide thumbs Also See for HI 8014:

Advertisement

Instruction Manual
HI 8014 - HI 8314
HI 8424 - HI 8915
HI 9214 - HI 931000
HI 9622
Portable
pH Meters
These Instruments are in
h t t p : / / w w w . h a n n a i n s t . c o m
Compliance with the CE Directives

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 8014 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments HI 8014

  • Page 1 Instruction Manual HI 8014 - HI 8314 HI 8424 - HI 8915 HI 9214 - HI 931000 HI 9622 Portable pH Meters These Instruments are in h t t p : / / w w w . h a n n a i n s t . c o m...
  • Page 2: Table Of Contents

    SPECIFICATIONS HI 8014 ............6 FUNCTIONAL DESCRIPTION HI 8314 ..........7 HI 8014, HI 8314, HI 8424, HI 8915, HI 931000 and HI 9622 SPECIFICATIONS HI 8314 ............8 are hand-held pH/mV meters. HI 9214 is a compact stick pH meter FUNCTIONAL DESCRIPTION HI 8915 ..........
  • Page 3: Functional Description Hi 8014

    4.01, pH 7.01 and pH 10.01. It can also be used for ISE FUNCTIONAL DESCRIPTION HI 8014 FUNCTIONAL DESCRIPTION HI 8014 FUNCTIONAL DESCRIPTION HI 8014 FUNCTIONAL DESCRIPTION HI 8014 FUNCTIONAL DESCRIPTION HI 8014 and ORP measurements with automatic range switching when readings exceed ±399.9 mV.
  • Page 4 S P E C I F I C A T I O N S H I 8 0 1 4 S P E C I F I C A T I O N S H I 8 0 1 4 HI 8014 Range 0.00 to 14.00...
  • Page 5 S P E C I F I C A T I O N S H I 8 3 1 4 S P E C I F I C A T I O N S H I 8 3 1 4 S P E C I F I C A T I O N S H I 8 3 1 4 S P E C I F I C A T I O N S H I 8 3 1 4 S P E C I F I C A T I O N S H I 8 3 1 4...
  • Page 6 S P E C I F I C A T I O N S H I 8 9 1 5 S P E C I F I C A T I O N S H I 8 9 1 5 S P E C I F I C A T I O N S H I 8 9 1 5 S P E C I F I C A T I O N S H I 8 9 1 5 S P E C I F I C A T I O N S H I 8 9 1 5...
  • Page 7 S P E C I F I C A T I O N S H I 9 2 1 4 S P E C I F I C A T I O N S H I 9 2 1 4 S P E C I F I C A T I O N S H I 9 2 1 4 S P E C I F I C A T I O N S H I 9 2 1 4 S P E C I F I C A T I O N S H I 9 2 1 4...
  • Page 8: Functional Description Hi 931000

    S P E C I F I C A T I O N S H I 8 4 2 4 S P E C I F I C A T I O N S H I 8 4 2 4 S P E C I F I C A T I O N S H I 8 4 2 4 S P E C I F I C A T I O N S H I 8 4 2 4 S P E C I F I C A T I O N S H I 8 4 2 4...
  • Page 9 S P E C I F I C A T I O N S H I 9 3 1 0 0 0 S P E C I F I C A T I O N S H I 9 3 1 0 0 0 S P E C I F I C A T I O N S H I 9 3 1 0 0 0 S P E C I F I C A T I O N S H I 9 3 1 0 0 0 S P E C I F I C A T I O N S H I 9 3 1 0 0 0...
  • Page 10: Operational Guide

    0.01 4 cm/1½") in a pH 7 or pH 4 buffer solution for a few minutes to 0.1mV reactivate it. For HI 8014: °C Connect the pH electrode to the BNC socket on the top of the meter. Accuracy ±0.01 Turn the meter on by turning the rotary knob to the °C, pH or mV...
  • Page 11 The display will show the pH value TAKING pH MEASUREMENTS compensated for temperature. For HI 8014: For HI 9214 and HI 931000: To take a pH measurement simply submerge After submerging the electrode into the sample the tip (4cm/1½") of the electrode and a...
  • Page 12 To enter the "mV" mode (ORP Oxidation Re- the pH reading. duction Potential) turn the instrument ON and °C select the mV mode (HI 8014, HI 8314 and HI 8915 only) or press the RANGE key until RANGE the display changes to mV (HI 8424 and...
  • Page 13 7.01 (HI 7007 or HI 8007) and pH 10.01 (HI 7010 or HI 8010) for alkaline measurements. If you need to calibrate HI 8014, HI 8314 and HI 8915 to NBS standards, use pH 6.86 (HI 7006 or HI 8006) instead of pH 7.01 and pH 9.18 (HI 7009 or HI 8009) instead of pH 10.01.
  • Page 14 °C mometer), e.g. 20.0°C. • Remove the protective cap from the elec- trode, rinse the tip of the electrode with • Set the rotary knob to °C to display the some pH 7.01 solution, then immerse the manual temperature setting. pH electrode and temperature probe into •...
  • Page 15 PROCEDURE FOR HI 8424 • Rinse the electrode and temperature probe before immersing them into pH 4.01 / • Switch the meter on after connecting the pH 10.01 calibration buffer, stir gently pH electrode and the temperature probe. and wait for thermal equilibrium. •...
  • Page 16 PROCEDURE FOR HI 9214: • Rinse the pH electrode with some pH 4.01 or 10.01 buffer solution (2 calibration • Turn the meter on. point). Note: the stability indicator "∆" blinks when • Dip the electrode into the pH 4.01 (or the electrode is not immersed in the pH 10.01) buffer solution.
  • Page 17 • If the ON/OFF key is pressed at this point, the calibration process • Press RANGE to display pH measurement. RANGE will end with only the offset (one-point) calibration. For best accuracy, however, it is recommended that a two-point calibra- tion is performed.
  • Page 18: Ph Values At Various Temperatures

    TEMPERATURES TEMPERATURES TEMPERATURES TEMPERATURES TEMPERATURES For manual temperature compensation during calibration (HI 8014, HI 8424 and HI 9622), please refer to the following chart. HI 7004 TEMP pH VALUES • When "CAL" disappears and the meter displays "CFM" press CFM °C...
  • Page 19: Temperature Calibration

    T R O U B L E S H O O T I N G G U I D E T R O U B L E S H O O T I N G G U I D E T R O U B L E S H O O T I N G G U I D E TEMPERATURE CALIBRATION TEMPERATURE CALIBRATION...
  • Page 20 SYMPTOMS PROBLEM SOLUTION ELECTRODE CONDITIONING ELECTRODE CONDITIONING ELECTRODE CONDITIONING ELECTRODE CONDITIONING ELECTRODE CONDITIONING For HI 8424 only A N D M A I N T E N A N C E A N D M A I N T E N A N C E A N D M A I N T E N A N C E A N D M A I N T E N A N C E A N D M A I N T E N A N C E...
  • Page 21 • For ORP Electrodes: polish the metal tip with a lightly upright for 1 hour. Follow the Storage Procedure above. abrasive paper (paying attention not to scratch the surface) and wash thoroughly with water. AmpHel is a registered Trademark of "Hanna Instruments" ®...
  • Page 22: Taking Redox Measurements

    TAKING REDOX MEASUREMENTS TAKING REDOX MEASUREMENTS - HI 9214 and HI 931000 display "Eb". HI 8014, HI 8314, HI 8424, HI 8915 and HI 9622 have the - HI 9622 displays "BAT" on the left side of capability to take ORP measurements.
  • Page 23: Display Codes Guide

    A C C E S S O R I E S A C C E S S O R I E S pH CALIBRATION SOLUTIONS HI 8014, HI 8314, HI 8424, HI 8915, HI 9214 AND HI 70004P pH 4.01 Buffer Sachets, 25 x 20 mL HI 9622 HI 7004M pH 4.01 Buffer Solution, 230 mL...
  • Page 24 HI 7074M Inorganic Cleaning Solution, 230 mL pH ELECTRODES HI 7074L Inorganic Cleaning Solution, 460 mL B = BNC CONNECTION PLUG + 1 m (3.3’) CABLE HI 7077M Oil & Fat Cleaning Solution, 230 mL HI 7077L Oil & Fat Cleaning Solution, 460 mL ELECTRODE CLEANING SOLUTIONS IN FDA S = SCREW-TYPE CONNECTOR APPROVED BOTTLE...
  • Page 25 HI 1330B / HI 1310S FC 100B Glass-body, semimicro, single junction, refillable, combination pH Plastic-body (Kynar ), double junction, refillable, combination pH ® electrode. Use: laboratory. electrode. Use: general purpose for food industry. 12 mm 5mm DIA 0.5" 0.2" 0.2" HI 1330 FC 100 120 mm...
  • Page 26 HI 7855/15 Extension cable 15 m (49.5') long Glass-body, refillable, combination platinum ORP electrode. Use: titration. OTHER ACCESSORIES HI 710001 Soft carrying case for HI 8014, HI 8314, HI 8424, 12 mm 0.5" HI 8915 and HI 9622 (formerly BORG)
  • Page 27: Electrode Application Reference Guide

    14. Galvanizing waste solution HI 1130B, HI 1912B, HI 1110S incurred. If the instrument is to be returned to Hanna Instruments, 15. Heavy-duty applications HI 1135B, HI 1115S 16.
  • Page 28: Other Products From Hanna

    CE DECLARATION OF CONFORMITY CE DECLARATION OF CONFORMITY CE DECLARATION OF CONFORMITY OTHER PRODUCTS FROM HANNA OTHER PRODUCTS FROM HANNA OTHER PRODUCTS FROM HANNA CE DECLARATION OF CONFORMITY CE DECLARATION OF CONFORMITY OTHER PRODUCTS FROM HANNA OTHER PRODUCTS FROM HANNA •...
  • Page 29 HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE LAB RECORDING WATER ANALYSIS HANDBOOK ENVIROCARE GENERAL CATALOG These and many others catalogs, handbooks and leaflets are avail- able from Hanna. To receive your free copy, contact your dealer or the nearest Hanna Customer Service Center.

This manual is also suitable for:

Hi 9622Hi 8424Hi 8314Hi 8915Hi 9214Hi 931000

Table of Contents