pH 1970i
Setting the
save interval:
ba75690e01
01/2007
4.4.2 Saving automatically
The save interval (Int 1) determines the chronological interval between
automatic save processes.
After the fixed interval has expired, the current data record is transmit-
ted to the storage and to the interface.
The default setting for the save interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min,
10 min, 15 min, 30 min, 60 min):
1
Press the <RUN/ENTER> key and hold it down.
2
Press the <STO> key. Int 1 appears on the display.
pH
S
I
n t
O
Sal
Time
Baud
d
O
F
Day.Month
No.
Year
Ident
Tref25
Tref20
Auto
Store
LoBat
nLF
Lin
Oxi
Cal
TEC
3
Set the required interval between the saving procedures with
<L> <M>.
4
Confirm with <RUN/ENTER>.
The number of free memory locations appears on the display.
pH
S
I
4 9
O
Sal
Time
Baud
F
R
E
Day.Month
No.
Year
Ident
Tref25
Tref20
Auto
Store
LoBat
nLF
Lin
Oxi
Cal
TEC
5
As soon as all of the 500 storage locations are occupied,
AutoStore is terminated (Int 1 = OFF).
If there are not enough storage locations available for your
measurements:
– Output and backup the data storage (see page 37) and
– clear the memory (see page 42).