About this Document This document describes the plug-in boards for the total pressure gauge control- ler TPG300, intended as a supplement to the documentation of the basic unit TPG300 (→ [1]). Validity This document applies to plug-in boards listed below...
Store modules in antistatic bags or contain- ers. Damage resulting from incorrect handling may lead to a revocation of the guarantee. INFICON accepts no responsibility nor warranty if the user or third parties • utilize the product not according to the defined use •...
2 Description 2.1 Pirani / Cold Cathode The CP300C9 board is a combined board containing one Pirani and one cold cathode measurement circuit. Measurement Board CP300C9 Pirani measurement circuit The Pirani measurement circuit has one gauge cable connector, two trimmer po- tentiometers and one analog signal output.
3.2 Interface and Relay Relay Number Board IF300P Contact type 1 change over contact each floating max. 50 V (ac) relative to other contacts and ground Characteristic data (ac) Switching voltage, max. 30 V (dc), 50 V (ac) Switching current, max. 1.5 A Switching power, max.
4 Installation General Use screened cables only (connect screen to barrel of connector). If both ends of the screen are connected to ground, compensating currents must be prevented (e.g. by connecting all involved units to a common power distributor). In a Profibus-DP installation (IF300P board), use the recommended special cable only (→...
4.3 Connecting the Cold Cathode Gauge Additional protective measures must be taken if certain processes in the vacuum system (e.g. flashovers) can cause hazardous voltages on the gauge terminals. Although the gauge cables are screened, they should not be routed in parallel to lines producing strong electrical noise.
Protection against unintentional separation of the triaxial connector. 4.4 Connecting the Each measurement circuit is equipped with an analog <OUTPUT> signal output. Matching connectors are included with Analog Signal each measurement board. 4.5 Connecting the Relays of the IF300P Pin Assignment The relay connector on the rear of the IF300P has the following pin assignment: Relay 1 IF 300P...
4.6 Connecting the Profibus-DP Interface to the IF300P In a Profibus-DP installation, use the recommended special cable only (→ [2]). Pin Assignment The interface connector on the rear of the IF300P has the following pin assign- ment: IF 300P Screen Do not connect RxD/TxD-P (B wire)
5 Adjustment Pirani measurement circuits are factory-adjusted to the standard gauge. Due to manufacturing tolerances, contamination of the gauges, and different cable lengths, deviations are inevitable. Cold cathode measurement circuits are factory-adjusted and require no readjust- ment. 5.1 Adjusting the Pirani Two trimmer potentiometers are available for compen- sating the gauge tolerances, gauge contamination, or Measurement Circuit...
No communication Pin 2 and 3 of the inter- Use the correct cable face cable not crossed Incorrect Baud rate Match Baud rate Incorrect data format Adhere to the format specified for the TPG300 (→ [1]) IG9972BEN (2017-09) Plug-In Boards.om...
Adhere to the standardized Profibus-DP data format (→ [2]) Cycle time >100 ms Incorrect firmware Firmware TPG300: 302-654 Firmware Profibus: V1.5 Incorrect Baud rate Set Baud rate to 19200 Baud At the controller TPG300, firmware 302-654 or higher IG9972BEN (2017-09) Plug-In Boards.om...
Page 16
Cold Cathode Measurment With SHV and Triax Lemo (standard plugs on each side). Cables (Test voltage: 6 kV (dc)) Length 80 °C 250 °C Pirani Measurement Measurement cables, high cables for IKR085 temperature version for IKR085 ITC220003 ITC230003 ITC220005 ITC230005 ITC220010 ITC230010 ITC220015...