Flush/Purge The Fluid Path - Biotek Synergy H1 Instructions For Use Manual

Hybrid multi-mode microplate reader for in vitro diagnostic use
Hide thumbs Also See for Synergy H1:
Table of Contents

Advertisement

Carefully lift the filter cube top from the cube.
4.
5.
Wearing linen or cloth gloves, grasp the mirror by its edges and lift it out of the
cube.
The mirrors are seated on a shelf in the bottom of the cube and are
not secured in place.
6.
Wet absorbent towels such as Kimwipes,
reagent-grade ethanol. Wear gloves or use enough toweling so that solvents do
not dissolve oils from your hands that can seep through the toweling onto the
coated surface.
7.
Drag the trailing edge of the ethanol-soaked Kimwipe across the surface of the
mirror, moving in a single direction. A minimal amount of pressure can be
applied while wiping. However, too much pressure will damage the mirror.
8.
Use the magnifying glass to inspect the surface; if debris is still visible, repeat
with a new Kimwipe.
9.
To replace the mirror, hold it by its edges, turn it so that its label is face-up and
readable, and place it on the shelf in the filter cube. See

Flush/Purge the Fluid Path

Applies only to Synergy H1 models with injectors.
At the end of each day that the dispense module is in use, flush the fluid path using the
Gen5 priming utility. Leave the fluid to soak overnight or over a weekend, and then purge
the fluid before using the instrument again.
Synergy H1
Flush/Purge the Fluid Path |
not
lens paper, with an anhydrous
Figure 11
Figure 11
.
31