If using the USB cable, the information can be found via the Windows
Control Panel, under Ports in the Hardware/Device Manager area of System
Properties (e.g., Serial Port (COM5)).
Select
Plug & Play
A Synergy H1 must be connected to the computer and turned on to
appear in the Available Plug & Play Readers list.
Click
6.
Test Comm
communication attempt is successful, return to Gen5's main screen.
If the communication attempt is
Is the reader connected to the power supply and turned on?
Is the communication cable firmly attached to both the reader and the
computer?
Did you select the correct Reader Type in Gen5?
Try a different Com port.
Did you install the USB driver software?
If you remain unable to get Gen5 and the reader to communicate with each
other, contact BioTek's Technical Assistance Center.

13: Run a System Test

1.
Turn on the incubator:
In Gen5, select
Click the
Enter a Requested temperature of at least 37°C and click
Return to Gen5's main screen.
Wait until the incubator temperature reaches the set point before
continuing.
2.
Select
System > Diagnostics > Run System Test
reader, select the
3.
When the test is complete, a dialog requesting additional information appears.
Enter the information and click
4.
The results report appears. Scroll down toward the bottom; the text should read
"SYSTEM TEST PASS."
5.
Turn off the incubator:
Synergy H1
.
. Gen5 attempts to communicate with the reader. If the
not
successful, try the following:
System > Instrument Control > Synergy H1
Pre-Heating
tab.
Synergy H1
and click
OK
13: Run a System Test |
. If prompted to select a
OK
.
.
11
.
On
.