Reference and Specification

6.1

Installation notes

The PACE1000 pressure indicator requires a set of connections with the exception of the
reference connection, this provides a reference to atmosphere for gauge sensors and
barometric sensors.
The gas density and type does not affect the accuracy of pressure measurement, assuming
that the UUT is at the same level (height) as the indicator or gas head correction is
accurately set.
Values of air density (kg m
dioxide by volume.
Air
pressure
(kPa)
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
Note: 100 kPa = 1 bar
[EN] English
-3
) for air of relative humidity 50% and containing 0.04% carbon
Table 6-1 Air Density Values
14
16
18
1.052
1.045
1.037
1.064
1.057
1.049
1.077
1.069
1.061
1.089
1.081
1.073
1.101
1.093
1.085
1.113
1.105
1.097
1.125
1.117
1.109
1.137
1.129
1.121
1.149
1.141
1.133
1.162
1.153
1.145
1.174
1.165
1.156
1.186
1.177
1.168
1.198
1.189
1.180
1.210
1.201
1.192
1.222
1.213
1.204
1.234
1.225
1.216
1.247
1.237
1.228
1.259
1.249
1.240
1.271
1.261
1.252
1.283
1.274
1.264
Air temperature (°C)
20
22
1.029
1.021
1.041
1.033
1.053
1.045
1.065
1.057
1.077
1.069
1.089
1.080
1.100
1.092
1.112
1.104
1.124
1.116
1.136
1.128
1.148
1.139
1.160
1.151
1.172
1.163
1.184
1.175
1.196
1.187
1.207
1.199
1.219
1.210
1.231
1.222
1.243
1.234
1.255
1.246
6 - 1
24
26
1.014
1,006
1.025
1.018
1.037
1.029
1.049
1.041
1.061
1.053
1.072
1.064
1.084
1.076
1.096
1.088
1.108
1.099
1.119
1.111
1.131
1.123
1.143
1.134
1.154
1.146
1.166
1.158
1.178
1.169
1.190
1.181
1.201
1.193
1.213
1.204
1.225
1.216
1.237
1.228
K0470 Revision A