PACE Pressure Indicators User Manual
Pneumatic connections
WARNING
PARALLEL THREADS MUST BE USED. FEMALE THREAD TYPE IS PARALLEL THREAD TO
ISO228/1 (DIN ISO228/1, JIS B0202) G1/8.
TAPERED THREADS NOT ALLOWED.
PACE rear
panel
PACE rear
panel
Connection
Input
Reference
For examples of adaptors (Ref: page 2 - 4).
Input Pressure (Figure 2)
1.
Make sure the user systems can be isolated and vented.
2.
Connect the Unit Under Test (UUT) to the output connection port.
Note: For instruments with NPT connections, use applicable bonded sealing as shown in figure
below.
[EN] English
PACE pressure port
PACE pressure port
ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
Parallel threads must be used.
Female thread type is parallel thread to
ISO228/1 (DIN ISO228/1, JIS B0202)
G1/8
Tapered threads NOT allowed.
2 - 3