6 Reference and Specification
RS232
Located on the rear panel an external RS232 connection requires the following:
Connector
Communications
Baud Rate power-up default 9600, no parity & handshake = xon/xoff
Baud rates selectable **
Parity
Flow control
Protocols
Heritage emulation
Terminator
** Selectable through the user interface.
RS232 Comms Range Setup
Note: The RS232 Comms Range values are set up by the user because of the hardware the
user desires to operate with the PACE 1000 equipment. This procedure only allows
supervisor to select these user pre-installed values. To initially define / change or
delete the RS232 Comms Range values, refer to (Ref: K0472 PACE SCPI Remote
Communications Manual).
K0470 Revision A
9-way 'D' female wired as per Table 2-1
RS232 point-to-point only (daisy chain is not supported)
2400, 4800, 9600, 19k2, 38k4, 57k6 & 115k2
None, Odd & Even
None, Hardware & xon/xoff
PACE SCPI
DPI 142/150, DPI 141
CR or LF or CR/LF
6 - 10
[EN] English