6.10 Options - GE K0470 User Manual

Pace indicators pressure automated calibration equipment
Table of Contents

Advertisement

6 Reference and Specification

6.10 Options

Option enable process
To enable soft options on a PACE instrument, use the following:
1.
Touch the top Measure area of the screen.
2.
Select Global Setup.
3.
Select Calibration.
4.
Enter a Calibration PIN 1234.
5.
Enter new option key xxxxxxxxxx (10 digits).
6.
After entry of this key PACE confirms the options have been enabled.
Note: Hardware options automatically enable post installation
Selects Analogue
Output Range
On/Off
Update rate of Analogue Output
option from the control module.
Front view of Analogue
Option PCB connector
Analogue Output Bandwidth = 0.5 x Update Rate (Hz)
K0470 Revision A
15-way female D connector
8
7
6
5
4
15
14
13
12
6 - 28
CONSULT HANDBOOK
ANALOGUE 1
15
22V to 26V
30V maximum with respect to
30V
max
chassis.
Rated output = 24V
To maintain PACE product safety, external
circuits connected to the instrument must meet
Safety Extra-Low Voltage (SELV) requirements.
3
2
1
30V
11
10
9
ANALOGUE 2
analogue connections
max
[EN] English

Advertisement

Table of Contents
loading

Table of Contents