Code The Meter - Bayer HealthCare Ascensia Elite XL User Manual

Bayer health care ascensia elite xl diabetes care system user guide
Table of Contents

Advertisement

Code the Meter whenever a new box of Test Strips
is opened.* This will match the Meter to the Test
Strip reactivity.
7

Code the Meter

HOW:
You will need the Code Strip found
in the new box of strips.
1. Tear open the packet found in
the new carton of Test Strips
and remove the Code Strip (save
the packet).
2. Insert the Code Strip into the
test slot as shown. A beep
sounds and a full display appears
briefly. Another beep sounds and
the Code Number (for example
F-5) appears in the display.
3. Remove the Code Strip and
store in packet.
*You do not need to code the Meter if
the new Code Strip number matches
the previous Code Strip number.
4
SECTION

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents