MESSAGE
"Lo" display.
CAUTION:
Glucose levels
below 50 mg/dL
(2.8 mmol/L)
may indicate a
potentially serious
medical condition.
"Hı" display.
CAUTION:
Glucose levels
above 250 mg/dL
(13.9 mmol/L)
may indicate
a potentially
serious medical
condition.
23
Meter and Test Strip Problems
CAUSE
1. The blood glucose
value may be
below 20 mg/dL
(1.1 mmol/L).
2. Test Strip not
filled completely.
3. Test Strip is
defective or has
deteriorated.
4. Wrong Code
Number (F#)
used.
1. The blood
glucose value
may be above
600 mg/dL
(33.3 mmol/L).
2. Same as #3
above.
SECTION
SOLUTION
1. Possible hypoglycemic con-
dition (low blood sugar).
Contact your physician
or diabetes counselor
about your condition.
2. Rerun test with new strip
and apply sufficient sample
to fill Test Strip. Hold test
end to sample until after
the "beep" .
3. Rerun test using a new,
unexpired Test Strip.
4. Use correct Code Strip pro-
vided with carton of Test
Strips being used.
1. Possible hyperglycemic con-
dition (high blood sugar).
Contact your physician
or diabetes counselor
about your condition.
2. Same as #3 above.
11