W-View Mode; Readout Method (Scan Mode) - Hamamatsu Photonics C11440-42U30 Instruction Manual

Orca-flash4.0 lt plus, digital cmos camera
Table of Contents

Advertisement

9-3

W-VIEW MODE

9-3-1

Readout method (scan mode)

The readout direction can be set for the top area and the bottom area independently in W-VIEW
Mode. (Figure 9-16, 17, 18, 19)
Figure 9-16 Top: top to center
/ Bottom: center to bottom
Figure 9-18 Top: top to center
/ Bottom: bottom to center
W-VIEW Mode supports Binning readout, Sub-array readout and Rapid rolling mode.
In sub-array mode, the vertical position of sub-array readout can be set independently in the top
and bottom areas.
The vertical height of sub-array readout is same in the both areas.
Please refer to 9-3-3 [Frame rate calculation] about the frame rate of each readout mode.
Note
Rapid Rolling mode is different from other modes.
C11440-42U30/C11440-42U31 Instruction manual_Ver.1.2
Figure 9-17 Top: center to top
Figure 9-19 Top: center to top
/ Bottom: bottom to center
/ Bottom: center to bottom
35

Advertisement

Table of Contents
loading

This manual is also suitable for:

C11440-42u31

Table of Contents