Penetration Panels Wall Openings; Penetration Panel Closet; Minimum Magnet Ceiling Height (Top View); Integrated Cooling Cabinet - GE SIGNA PIONEER Installation Manual

Table of Contents

Advertisement

PENETRATION PANELS WALL OPENINGS

INTEGRATED COOLING CABINET

Front View
Equipment room side
Mesh shield
frame area
59
824
[2.32 in]
[32.44 in
]
6140
156
[
]
241.74 in
6.15 in
942
[37.1 in]
INTEGRATED COOLING CABINET
Top View
Magnet room
Equipment
room
Integrated System Cabinet and Integrated Cooling Cabinet must be covered on both sides for safety.
Typical
INTEGRATED SYSTEM CABINET
Front View
1107.2
[43.59 in
]
59
[2.32 in]
1226
[48.27 in]
EQUIPMENT ROOM FLOOR
INTEGRATED SYSTEM CABINET
Top View
Finished
wall
SIGNA PIONEER
59
[2.32 in]
A
Finished wall
A:
300 mm [11.81in] side space for excess cable sorage. Left side space and right side space can be swapped.
NOTE:All RF mesh Assy Shipping brackets should be kept on frame until after installed and RF test is completed.
An enclosure (i.e. closet) must be provided to restrict access to the PEN panels and for storage of excess interconnections.
The PEN closet must have a mechanical locking mechanism to restrict access to the PEN panels
The PEN closet must maintain the minimum service area outside the 200 Gauss in the magnet room.
PEN closet must allow free air exchange of
system blowers. Airflow may be achieved through door louvers or other openings in the PEN closet that meet all other
PEN closet requirements

MINIMUM MAGNET CEILING HEIGHT (TOP VIEW)

Shaded area within solid lines indicates floor to ceiling minimum height of 2500 mm [98.42in]. If ceiling is between
2500 mm [98.42in] and 2667 mm [105in], flexible main lead extension for low ceiling height is required for
ramping the magnet. Contact the PMI for preparation.
RF Shield
SCALE 1:50
EN-MRI-TYP-SIGNA_PIONEER-WEB.DWG

PENETRATION PANEL CLOSET

570
570
430
430
[22.4 in]
[22.4 in]
[16.9 in]
[16.9 in]
122
[4.8 in]
2600
[102.4 in]
between the Magnet room and PEN closet for MR
2769
[109 in]
MAGNET
GEOMETRIC
ISOCENTER
FRONT OF MAGNET
1372
[54 in]
Rev
A
Date
23/Jul/2018
minimum closet
inner dimensions
outline
ICC/ISC
200
[7.9 in]
50
[2 in]
RF
shield
400
[15.7 in]
1372
[54 in]
RF shielding
07
/24