GE SIGNA PIONEER Installation Manual page 4

Table of Contents

Advertisement

Moving metal sensitivity line for
CARS, MINIVANS, PICKUP
TRUCKS, AND AMBULANCES.
NOTE: Ferrrous objects must not move into or inside of the
moving metal sensistivity line during scans.
Moving metal sensitivity line for
BUSES AND TRUCKS (DUMP,
TRACTOR TRAILER, UTILITY, FIRE TRUCKS)
18'-8"
9'-4"
14
4
29
16
1
2
27
MAGNET ROOM
(considering RF
shield walls)
19
22
13
32
15
8
9
26
10
CONTROL ROOM
13'-2"
Typical
SIGNA PIONEER
6'-6"
6
28
5
28
11
23
23
18
30
7
20
3
EQUIPMENT
ROOM
25
24
12
17
EN-MRI-TYP-SIGNA_PIONEER-WEB.DWG
GE Supplied
A
GE Supplied/contractor installed
B
200 Gauss
100, 50, 30, 10 Gauss
BY
ITEM
DESCRIPTION
A
1
3T Magnet
A
2
Patient table
Magnet rundown unit
A
3
Phantom set storage cabinet
A
4
A
5
Integrated Cooling Cabinet
A
6
Integrated System Cabinet
Magnet monitor
A
7
Operator console computer
A
8
Operator workspace
A
9
A
10
Pneumatic patient alert
A
11
MR Elastography
700 va partial UPS
A
12
A
13
CCTV Monitor
A
14
CCTV Camera
A
15
Music system
D
16
Injector on pedestal
Injector control
D
17
Injector power supply
D
18
D
19
Remote magnet rundown unit
D
20
Main disconnect panel
Dimplex chiller
D
21
Remote graphic display
D
22
C
23
Breaker boxes for Integrated Cooling & Integrated Systems Cabinets
Minimum opening for equipment delivery is 40 in. w x 82 in. h, contingent on a 72 in.
C
24
corridor width
Minimum opening for is 43 in. w x 82 in. h, contingent on a 96 in. corridor width
C
25
Counter top for equipment- provide grommeted openings as required to route cables
C
26
C
27
Base cabinet for storage of: surface coils, patient positioning pads, phantoms, etc.
C
28
Louvered doors - refer to preinstall for requirements
Magnet access 9'-0"x10'-0"
C
29
C
30
Shelf
RF SHIELD - 100 dB ATTENUATION
21
LOCATED ELSEWHERE
Finished floor to slab height
Recommended finished ceiling height
Rev
A
Date
LEGEND
C
Customer/contractor supplied and installed
D
Available from GE
5 Gauss
3, 1, 0.5 Gauss
MAX
MAX HEAT
WEIGHT
HEAT
OUTPUT
(lbs)
OUTPUT
(btu)
(W)
8189
16500
2400
-
572
-
-
7
-
-
350
-
3412
1034
1000
24232
2758
7100
819
10
240
4947
122.80
1450
-
20
-
-
0.5
-
480
53.4
0.14
-
26
-
-
8.27
-
-
0.35
-
-
-
-
-
94
-
675
17
198
660
6
193
-
7
-
901
130
264
130000
1840
36000
-
-
-
Exam room height
23/Jul/2018
A2 - Equipment Layout
WEIGHT
(kg)
7500
259.50
3.2
136
469
1251
4.5
55.70
9
0.2
24.22
12
3.75
0.16
-
43
8
3
3.20
59
835
-
TBD
8'-9"
04
/24

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents