Advertisement

Analog Output 1
DC Power input
(15 – 30 VDC)
Active or Passive
TB2
AC Power input
(100 – 264 VAC)
TB1
DC INPUT
LOOP1/HART
PULSE/
AC
INPUT
ALARM
OUTPUT
100-264 VAC
A-
-
+
50/
60Hz
P-
P+
A+
P-
F2 TEST
TB2
TB3
P-
P+
LOOP2
TB1
F1
OUTPUT
TEST
F1
TB5
Control
Room
Display
Current
Meter
Figure 12
54-631 THERMATEL Model TA2 Transmitter
2.6
Output
connections
TB3
A-
P+ A+
D6
R1
TB4
POWER
Junction

2.6.1 Connection

> 250 Ω
R
L
-
+
Power
Supply

2.6.2 HART Revision Table

HART Version
Dev V2 DD V1

2.6.3 HART Display Menu

Configuration Using HART

A HART
®
(Highway Addressable Remote Transducer)
remote unit, such as a HART communicator, can be used
to provide a communication link to the TA2.
When connected to the control loop, the same system
measurement readings shown on the transmitter are
displayed on the communicator. The communicator can
also be used to configure the transmitter.
To confirm HART handheld communications, attach the
unit as shown in figure 12. If the communicator reads
GENERIC on the first two lines, then the
HART handheld does not contain the current DDs
(Device Descriptions) for the TA2 transmitter.
Contact your local HART Service Center.
The TA2 can be used with a HART remote unit to provide
communication of the process variables. The communicator
can also be used for configuration of the transmitter.
The dynamic variables which can be transmitted over
HART are Flow, Mass, Temperature, and Totalized Flow.
The Primary Variable is Flow or Mass.
Virtually all configuration and diagnostics available via the
user interface with the keypad and display are accessible
via HART. See System onfiguration Menu, Section 2.6.3.
A HART device can be connected to either the active or
passive 4-20 mA output loop from AO1.
HCF Release Date Compatible with TA2 Software
January 2015
When connected, the top line of each menu displays the
model (TA2) and its tag number or address.
Open the TA2 online menu tree by pressing alphanumeric
key 1 to enter the Device Setup menu.
®
Version 2.1a0