Introduction

Compatibility

Inspecting the Probe

Keysight N2790A Differential Voltage Probe
User Guide
The N2790A differential probe allows you to safely measure high-voltage
floating circuits with the oscilloscope grounded. It is ideal for many
applications such as motor speed controls, power supply designs, and
electronic high-power converters. The N2790A offers sufficient dynamic
range and bandwidth for your application to make the floating
measurements found in high-speed power electronic circuits safely and
accurately.
With 100 MHz bandwidth, switchable attenuation of 50:1 and 500:1, and a
maximum differential voltage input of 1400V (DC + peak AC), it provides
the versatility for a broad range of applications including high-voltage
circuits. The probe is powered by the auto-probe interface.
The N2790A differential probe is compatible with the following Keysight
oscilloscopes:
InfiniiVision Series 5000, 6000 (except 100 MHz), and 7000 with version
5.2 software or later
Infiniium 9000 (with version 2.00 software or later), 8000 (with version
5.61 software or later), and 54830 Series (with version 5.61 software or
later)
Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the
contents of the shipment have been checked for completeness and the
probe has been checked mechanically and electrically.
Check the accessories.
If the contents are incomplete or damaged, notify your Keysight
Technologies Sales Office.
Inspect the instrument.
If there is mechanical damage or defect, or if the probe does not
operate properly or pass calibration tests, notify your Keysight
Technologies Sales Office.